Impact of the fixed gantry angle approximation on dosimetric accuracy for helical tomotherapy plans.

G Samuel J Tudor1, Simon J Thomas.   

Abstract

PURPOSE: The purpose of the work was to determine the accuracy of the dose calculation of off-axis, small target helical tomotherapy treatments using 51 calculation angles, and to determine the increase in calculation angles required to improve the accuracy to acceptable standards.
METHODS: A previously described dose calculation program [S. J. Thomas, K. R. Eyre, G. S. J. Tudor, and J. Fairfoul, "Dose calculation software for helical tomotherapy, utilizing patient CT data to calculate an independent three-dimensional dose cube," Med. Phys. 39, 160-167 (2012)] was modified to permit decomposition of each projection into several subprojections, allowing more accurate modeling of the temporal distribution of fluence within each beamlet. Four plans of small off-axis spherical targets were recalculated several times with different numbers of subprojections, with the minimum dose to 95% of the PTV (D(95%)) and the minimum dose to 2% of the PTV (D(2%)) calculated for each, in order to determine the minimum number of subprojections required for accurate dose statistics. A further nine plans were used to determine the effect on conventional calculation accuracy of varying target size, target position, modulation factor, and pitch. For this analysis, the mean dose and equivalent uniform dose were considered in addition to D(95%) and D(2%).
RESULTS: The differences between calculations made using the 51 angle approximation and using the closest approximation to real treatment delivery were notable, with up to 11.0% overestimate of D(95%) for the cases studied. A previously unreported underestimate of dose to parts of the PTV was observed due to this effect, with D(2%) being underestimated by up to 3.3%. The effect is dependent on target size, position, modulation factor, and the angular distribution of fluence within the sinogram but not pitch. Decomposing each projection into three subprojections left differences in dose statistics that were of reduced magnitude but still appreciable. The effect of increasing the number of subprojections beyond five had little effect.
CONCLUSIONS: When applied to small, off-axis targets, the limitations of the 51 calculation angle model can substantially affect the veracity of PTV dose statistics, including both underestimation and overestimation of dose depending on position within the PTV. Increasing the number of calculation angles by a factor of 5 reduces the effect to insignificant levels. While the latest release of TomoTherapy planning software will ameliorate the problem, the studied effect is best avoided by positioning small targets near to the bore center. Where this is not possible, it is recommended to ensure a high actual modulation factor and to use an unscaled delivery for patient-specific quality assurance.

Entities:  

Mesh:

Year:  2013        PMID: 23298081     DOI: 10.1118/1.4769120

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Using 4D dose accumulation to calculate organ-at-risk dose deviations from motion-synchronized liver and lung tomotherapy treatments.

Authors:  William S Ferris; Edward H Chao; Jennifer B Smilowitz; Randall J Kimple; John E Bayouth; Wesley S Culberson
Journal:  J Appl Clin Med Phys       Date:  2022-04-29       Impact factor: 2.243

2.  Recalculation of dose for each fraction of treatment on TomoTherapy.

Authors:  Simon J Thomas; Marina Romanchikova; Karl Harrison; Michael A Parker; Amy M Bates; Jessica E Scaife; Michael P F Sutcliffe; Neil G Burnet
Journal:  Br J Radiol       Date:  2016-01-05       Impact factor: 3.039

3.  Implementation of TomoEDGE in the independent dose calculator CheckTomo.

Authors:  Mathieu Schopfer; Simon J Thomas; G Samuel J Tudor; Jean Bourhis; François Bochud; Raphaël Moeckli
Journal:  J Appl Clin Med Phys       Date:  2017-03-06       Impact factor: 2.102

4.  Measurement-guided volumetric dose reconstruction for helical tomotherapy.

Authors:  Cassandra Stambaugh; Benjamin Nelms; Theresa Wolf; Richard Mueller; Mark Geurts; Daniel Opp; Eduardo Moros; Geoffrey Zhang; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

5.  Comparison of dose calculations between pencil-beam and Monte Carlo algorithms of the iPlan RT in arc therapy using a homogenous phantom with 3DVH software.

Authors:  Jin Ho Song; Hun-Joo Shin; Chul Seung Kay; Soo-Min Chae; Seok Hyun Son
Journal:  Radiat Oncol       Date:  2013-12-05       Impact factor: 3.481

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.