| Literature DB >> 2329396 |
C E Crofts1, A Trowbridge, T Maung Aung, I M Brook.
Abstract
The use of porcine rib pairs as an in vitro analog for the edentulous mandible is described. Using this model, the relative degree of fixation achieved with a screw plate (Champy) and a paraskeletal clamp plate (Mennen) has been evaluated. The fractured plated ribs failed at significantly lower bending forces than the non-sectioned controls. However, no statistically significant difference in force at failure between the two different methods of plate fixation could be demonstrated.Entities:
Mesh:
Year: 1990 PMID: 2329396 DOI: 10.1016/0278-2391(90)90231-p
Source DB: PubMed Journal: J Oral Maxillofac Surg ISSN: 0278-2391 Impact factor: 1.895