Literature DB >> 23292554

Benchtop testing and comparisons among three types of through-the-scope endoscopic clipping devices.

Sumanth R Daram1, Shou-Jiang Tang, Ruonan Wu, S D Filip To.   

Abstract

BACKGROUND: Through-the-scope (TTS) endoscopic clipping devices are widely used. No benchtop testing or direct comparisons of these endoclips have been performed to show their rotational ability and inherent mechanical strengths during closure and after deployment. This study aimed to provide benchtop data that can be used to guide clinical applications and to promote future device research and development.
METHODS: Benchtop testing and comparisons were performed for three groups of TTS clips: QuickClip2 long, resolution, and instinct clips. The main outcome measurements were device-in-endoscope retroflection angles (DIERA), opening strength, "snapping" force of acute clip closure, and neoprene pulling strength.
RESULTS: The achievable gastroscope DIERA was 10° for QuickClip2, 3° for the resolution clip, and 10° for the instinct clip. The QuickClip and the Instinct clip rotated almost equally well under all endoscope configurations, including endoscopic retrograde cholangiopancreatography (ERCP). With or without a sheath, the resolution clip lacked the ability to rotate. During clip opening force testing (the amount of force required to force open the jaws of a deployed clip by 3.2 mm; 3.2 mm was chosen due to the standard dimension of the gauge used for the measurement), the Instinct clips were the strongest. For the Instinct clips, an opening force of 404 ± 124 g was needed to open the closed clip, and an additional 386 ± 133 g was required to open the clip jaws to 3.2 mm. In terms of snapping force during acute closure and neoprene pulling strength, the instinct and resolution clips performed almost equally. The limitations of the study were the benchtop testing and the finite sample size for closing and pulling strength comparisons.
CONCLUSIONS: The QuickClip2 and the Instinct clip rotate equally well under different endoscope configurations. The resolution clips lack rotational ability. The instinct clips are stronger mechanically than the other two TTS clips. Stronger clips are perhaps associated with higher therapeutic efficacy and retention rates.

Entities:  

Mesh:

Year:  2013        PMID: 23292554     DOI: 10.1007/s00464-012-2679-4

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  15 in total

Review 1.  Endoscopic full-thickness GI wall resection: current status.

Authors:  Yael Kopelman; Peter D Siersema; Amol Bapaye; Doron Kopelman
Journal:  Gastrointest Endosc       Date:  2012-01       Impact factor: 9.427

2.  "Hemodynamic efficacy" of two endoscopic clip devices used in the treatment of bleeding vessels, tested in an experimental setting using the compact Erlangen Active Simulator for Interventional Endoscopy (compactEASIE) training model.

Authors:  J Maiss; C Dumser; Y Zopf; A Naegel; N Krauss; J Hochberger; K Matthes; E G Hahn; D Schwab
Journal:  Endoscopy       Date:  2006-06       Impact factor: 10.093

Review 3.  Endoscopic mechanical hemostasis of GI arterial bleeding (with videos).

Authors:  Gottumukkala S Raju; Tonya Kaltenbach; Roy Soetikno
Journal:  Gastrointest Endosc       Date:  2007-10       Impact factor: 9.427

Review 4.  Endoscopic hemostatic devices.

Authors:  Jason D Conway; Douglas G Adler; David L Diehl; Francis A Farraye; Sergey V Kantsevoy; Vivek Kaul; Sripathi R Kethu; Richard S Kwon; Petar Mamula; Sarah A Rodriguez; William M Tierney
Journal:  Gastrointest Endosc       Date:  2009-05       Impact factor: 9.427

5.  A randomised controlled comparison of injection, thermal, and mechanical endoscopic methods of haemostasis on mesenteric vessels.

Authors:  C C Hepworth; S S Kadirkamanathan; F Gong; C P Swain
Journal:  Gut       Date:  1998-04       Impact factor: 23.059

6.  A comparative study of intracranial aneurysm clips: closing and opening forces and physical endurance.

Authors:  K Ooka; M Shibuya; Y Suzuki
Journal:  Neurosurgery       Date:  1997-02       Impact factor: 4.654

7.  Randomized controlled study of 3 different types of hemoclips for hemostasis of bleeding canine acute gastric ulcers.

Authors:  Dennis M Jensen; Gustavo A Machicado; Ken Hirabayashi
Journal:  Gastrointest Endosc       Date:  2006-11       Impact factor: 9.427

8.  Prospective, randomized comparison of 3 different hemoclips for the treatment of acute upper GI hemorrhage in an established experimental setting.

Authors:  Masayuki Kato; Yunho Jung; Mark A Gromski; Ram Chuttani; Kai Matthes
Journal:  Gastrointest Endosc       Date:  2012-01       Impact factor: 9.427

9.  Titanium aneurysm clips: mechanical characteristics and clinical trial.

Authors:  T Nagatani; M Shibuya; K Ooka; Y Suzuki; M Takayasu; J Yoshida
Journal:  Neurol Med Chir (Tokyo)       Date:  1998       Impact factor: 1.742

10.  Comparative study of endoscopic clips: duration of attachment at the site of clip application.

Authors:  Eun Ji Shin; Chung-Wang Ko; Priscilla Magno; Samuel A Giday; John O Clarke; Jonathan M Buscaglia; Gevorg Sedrakyan; Sanjay B Jagannath; Anthony N Kalloo; Sergey V Kantsevoy
Journal:  Gastrointest Endosc       Date:  2007-10       Impact factor: 9.427

View more
  9 in total

1.  Evaluating long-term attachment of a novel endoclip in porcine stomachs: a prospective study of initial deployment success and clip retention rates at different regions of the stomachs.

Authors:  Bo Yan; Rui-Hua Shi; Ya-Dong Feng; Zhen-Hai Di
Journal:  Surg Endosc       Date:  2015-06-20       Impact factor: 4.584

2.  Not all through-the-scope endoscopic clipping devices are born equal: some can be rotated while others cannot.

Authors:  Shou-jiang Tang
Journal:  Surg Endosc       Date:  2013-04-30       Impact factor: 4.584

3.  Can all through-the-scope endoscopic clipping devices be rotated? Yes, they can.

Authors:  Payal Saxena; Mouen A Khashab
Journal:  Surg Endosc       Date:  2013-04-30       Impact factor: 4.584

4.  Deployment of a Short, Single-Opening Endoscopic Clip Versus a Long, Reopening Endoscopic Clip in Clinical Practice.

Authors:  Eric Wee; Mathew Philip Sachin; Uthamanand Chinnappa; Su Chang; Cherng Hann Benjamin Yip
Journal:  Dig Dis Sci       Date:  2015-03-31       Impact factor: 3.199

Review 5.  New and emerging endoscopic haemostasis techniques.

Authors:  Rebecca Palmer; Barbara Braden
Journal:  Frontline Gastroenterol       Date:  2015-01-23

6.  Immediate technical and delayed clinical outcome of fistula closure using an over-the-scope clip device.

Authors:  Ryan Law; Louis M Wong Kee Song; Shayan Irani; Todd H Baron
Journal:  Surg Endosc       Date:  2014-10-03       Impact factor: 4.584

Review 7.  Endoscopic hemostasis state of the art - Nonvariceal bleeding.

Authors:  Stefan Karl Goelder; Juliane Brueckner; Helmut Messmann
Journal:  World J Gastrointest Endosc       Date:  2016-02-25

8.  Which clip? A prospective comparative study of retention rates of endoscopic clips on normal mucosa and ulcers in a porcine model.

Authors:  Payal Saxena; Eun Ji-Shin; Yamile Haito-Chavez; Ali K Valeshabad; Venkata Akshintala; Gerard Aguila; Vivek Kumbhari; Dawn S Ruben; Anne-Marie Lennon; Vikesh Singh; Marcia Canto; Anthony Kalloo; Mouen A Khashab
Journal:  Saudi J Gastroenterol       Date:  2014 Nov-Dec       Impact factor: 2.485

9.  Cap-assisted hemoclip application with forward-viewing endoscope for hemorrhage induced by endoscopic sphincterotomy: a prospective case series study.

Authors:  Feng Liu; Guang-Yong Wang; Zhao-Shen Li
Journal:  BMC Gastroenterol       Date:  2015-10-15       Impact factor: 3.067

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.