| Literature DB >> 23289768 |
Carlo Luca Romanò1, Roberto D'Anchise, Marco Calamita, Giovanni Manzi, Delia Romanò, Valerio Sansone.
Abstract
BACKGROUND: Diagnosis of peri-prosthetic infection remains challenging, often requiring a combination of different tests.Entities:
Mesh:
Year: 2013 PMID: 23289768 PMCID: PMC3565984 DOI: 10.1186/1471-2474-14-7
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Figure 1Right knee infected prosthesis. Clinical aspect (A) and thermogram (B), with maximum (Max) and average (Med) recorded temperature and temperature distribution (C).
Figure 2Aseptic failed left knee prosthesis. Clinical aspect (A) and thermogram (B), with maximum (Max) and average (Med) recorded temperature and temperature distribution (C).
Isolated pathogens the septic group of patients (N = 36)
| CNS | 10 | 27.8 |
| MRSE | 3 | 8.3 |
| MSSA | 8 | 22.2 |
| MRSA | 5 | 13.9 |
| Pseudomonas aeruginosa | 4 | 11.1 |
| Enterococcus spp. | 3 | 8.3 |
| E. coli | 1 | 2.7 |
| Pasteurella multocida | 1 | 2.7 |
| No isolates | 8 | 22.2 |
CNS: coagulase-negative staphylococci; MRSE: Methicillin-Resistant Staph. Epidermidis; MSSA: Methicillin-Sensitive Staph. Aureus; MRSA: Methicillin-Resistant Staph. Aureus (more than one microrganism may be isolated form a given patient).
Pre-operative clinical assessment of the reviewed patients (N = 70) (more than one sign may be present in a patient)
| Swelling | 26 (72%) | 18 (53%) |
| Local redness | 22 (61%) | 5 (15%) |
| Local warmth (clinical perception) | 22 (61%) | 7 (20%) |
| Stiffness | 10 (28%) | 11 (32%) |
| Draining fistula | 7 (19%) | 0 |
| Fever | 3 (8%) | 0 |
Pre-operative laboratory tests (N = 70)
| | |||||
|---|---|---|---|---|---|
| C-RP (mg/L) | 38.4 ± 41.2 | 2 - 135 | 6.4 ± 7.1 | 0 – 22 | 0.0001 |
| ESR (mm/hr) | 52 ± 43 | 15 - 160 | 11 ± 18 | 4 – 46 | 0.0001 |
| WBC / ml | 7400 ± 2200 | 4200 - 13700 | 6500 ± 2200 | 4200 – 13700 | 0.09 |
| IL-6 (pg/ml) | 8.8 ± 3.8 | 4.9 – 12.2 | 2.9 ± 0.4 | 2.8 - 3.2 | 0.0001 |
Figure 3Mean recorded temperatures of septic (A) and aseptic (B) failed knee prosthesis. HS: Hottest Sport. SSA: Surgical Site Area. Affected: painful knee. N-Aff: sound knee. * P < 0.0001.
Figure 4Receiver Operating Curve (ROC) for temperature threshold assessment for Surgical Site Area (A) and Hottest Spot (B). Dotted lines indicate the temperature at which the highest value of sensitivity + specificity was calculated.