Appropriate collimators in a small animal SPECT scanner with CZT detector.

Yusuke Higaki1, Masato Kobayashi, Tomoya Uehara, Hirofumi Hanaoka, Yasushi Arano, Keiichi Kawai.   

Abstract

OBJECTIVE: Almost all small animal SPECT is performed with pinhole collimators (PH), including single-PH (SPH) and multi-PH (MPH). In the clinical study, not only PH but also parallel-hole collimator (PAH) is often used in planar and SPECT imaging. However, there have been no comparative studies on image quality with various collimators on the small animal imaging. This study compared the basic characteristics of PH and PAH in small animal imaging.
METHODS: Performance of planar and SPECT images was evaluated using (99m)TcO4(-) and SPH, MPH and PAH with low energy and high resolution on the SPECT/CT scanner FX3200. We measured sensitivity, resolution, concentration linearity and uniformity. Planar imaging of mice with (99m)Tc-labeled mercaptoacetyltriglycine ((99m)Tc-MAG3) was performed using SPH and PAH. SPECT imaging with (99m)Tc-methylene diphosphonate ((99m)Tc-MDP) was performed using all collimators.
RESULTS: With SPH, MPH and PAH, sensitivity was 43.5, 211.2 and 926.5 cps/MBq, respectively, and spatial resolution was 0.60/0.56, non/0.96, 5.20/5.34 mm full-width half maximum (planar/SPECT), respectively. There were marked correlations between the radioactivity counts on images and radioactivity with all collimators. Values of % standard deviation on planar imaging showed small differences between the SPH and PAH, while the values were the smallest on SPECT imaging with MPH. On imaging of mice, SPH yielded high-quality (99m)Tc-MAG3-planar images when compared with PAH. MPH yielded sharper (99m)Tc-MDP-SPECT images than SPH and PAH.
CONCLUSIONS: The characteristics of PH and PAH differed on small animal imaging. Although sensitivity was higher with PAH, PH showed higher resolution. Among the PH collimators, SPH was more appropriate for planar imaging, and MPH was more suitable for SPECT imaging in a small animal imaging scanner with CZT detector.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23288435     DOI: 10.1007/s12149-012-0681-5

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  2 in total

1.  Impact of injection dose, post-reconstruction filtering, and collimator choice on image quality of myocardial perfusion SPECT using cadmium-zinc telluride detectors in the rat.

Authors:  Asuka Mizutani; Ichiro Matsunari; Masato Kobayashi; Kodai Nishi; Wataru Fujita; Yoshiharu Miyazaki; Stephan G Nekolla; Keiichi Kawai
Journal:  EJNMMI Phys       Date:  2015-03-11

2.  Capabilities of multi-pinhole SPECT with two stationary detectors for in vivo rat imaging.

Authors:  Jan P Janssen; Jan V Hoffmann; Takayuki Kanno; Naoko Nose; Jan-Peter Grunz; Masahisa Onoguchi; Xinyu Chen; Constantin Lapa; Andreas K Buck; Takahiro Higuchi
Journal:  Sci Rep       Date:  2020-10-29       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.