Twenty-four-hour measurement of IOP in rabbits using rebound tonometer.

Xiaogang Wang1, Jing Dong, Qiang Wu.   

Abstract

PURPOSE: To observe the Intraocular pressure (IOP) measurements over a 24-h period in New Zealand white rabbits.
METHODS: Thirty-four normal New Zealand white rabbits (68 eyes) were included in this study. The IOP values were determined with a rebound tonometer (RBT, Icare VET). Measurements were always obtained in the right eye first.
RESULTS: The data showed that there was no significant difference in IOP between measurements, which indicated that the RBT showed good repeatability in this study (P = 0.920). There was no statistically significant difference between genders (P = 0.943). The IOP in right eyes was a little higher than left eyes (F = 45.96, P < 0.001). Comparison of the true IOP and RBT results over the entire pressure range (8-100 mmHg) revealed a statistically significant difference between them (P < 0.001). The mean difference between the RBT and true IOP (RBT-True IOP) was -22.65 ± 9.52 mmHg with 95% confidence interval (-27.1, -18.2 mmHg). In the light period (7 AM to 7 PM), the IOP was a little lower than during the dark period (8 PM to 6 AM); 9 AM and 2 PM were the lowest measurements (10 mmHg, 9.625 mmHg), and 10 AM (14.125 mmHg)was the highest measurement during the 24-h period.
CONCLUSIONS: The IOP obtained by RBT underestimated the true IOP but the RBT had a good repeatability in IOP measurements. In the light period (7 AM to 7 PM), the IOP was a little lower than during the dark period (8 PM to 6 AM), which may be beneficial to the intervention therapy observation of glaucoma research in rabbits.
© 2012 American College of Veterinary Ophthalmologists.

Entities:  

Keywords:  24-h; intraocular pressure; rabbits; rebound tonometer

Mesh:

Year:  2013        PMID: 23279703     DOI: 10.1111/vop.12020

Source DB:  PubMed          Journal:  Vet Ophthalmol        ISSN: 1463-5216            Impact factor:   1.644


  5 in total

1.  Twenty-four-Hour Measurement of Intraocular Pressure in Guinea Pigs (Cavia porcellus).

Authors:  Maneli Ansari-Mood; Seyed Mehdi-Rajaei; Reza Sadjadi; Masoud Selk-Ghaffari; David L Williams
Journal:  J Am Assoc Lab Anim Sci       Date:  2016-01       Impact factor: 1.232

2.  The impacts of individual and combined exposure to cadmium and lead on intraocular pressure, electroretinography, and residual changes in the rabbit eyes.

Authors:  Yasmina M Abd-Elhakim; Khlood M El Bohi; Nabela I El Sharkawy; Manar A Ghali; Sarah Haseeb
Journal:  Environ Sci Pollut Res Int       Date:  2019-09-13       Impact factor: 4.223

3.  Repeatability, reproducibility and agreement of intraocular pressure measurement in rabbits by the TonoVet and Tono-Pen.

Authors:  Di Ma; Chong-Bo Chen; Jiajian Liang; Zhihao Lu; Haoyu Chen; Mingzhi Zhang
Journal:  Sci Rep       Date:  2016-10-12       Impact factor: 4.379

4.  Comparison of two prototypes of a magnetically adjustable glaucoma implant in rabbits.

Authors:  Birthe Schwerk; Lisa Harder; Claudia Windhövel; Marion Hewicker-Trautwein; Anna Wagner; Jan-Peter Bach; Lena Carolin Voigt; Ulf Hinze; Boris Chichkov; Heinz Haferkamp; Holger Lubatschowski; Stephan Nikolic; Ingo Nolte
Journal:  PLoS One       Date:  2019-04-11       Impact factor: 3.240

Review 5.  Species Differences in the Nutrition of Retinal Ganglion Cells among Mammals Frequently Used as Animal Models.

Authors:  Christian Albrecht May
Journal:  Cells       Date:  2019-10-14       Impact factor: 6.600

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.