Literature DB >> 23278024

High performance imaging streak camera for the National Ignition Facility.

Y P Opachich1, D H Kalantar, A G MacPhee, J P Holder, J R Kimbrough, P M Bell, D K Bradley, B Hatch, G Brienza-Larsen, C Brown, C G Brown, D Browning, M Charest, E L Dewald, M Griffin, B Guidry, M J Haugh, D G Hicks, D Homoelle, J J Lee, A J Mackinnon, A Mead, N Palmer, B H Perfect, J S Ross, C Silbernagel, O Landen.   

Abstract

An x-ray streak camera platform has been characterized and implemented for use at the National Ignition Facility. The camera has been modified to meet the experiment requirements of the National Ignition Campaign and to perform reliably in conditions that produce high electromagnetic interference. A train of temporal ultra-violet timing markers has been added to the diagnostic in order to calibrate the temporal axis of the instrument and the detector efficiency of the streak camera was improved by using a CsI photocathode. The performance of the streak camera has been characterized and is summarized in this paper. The detector efficiency and cathode measurements are also presented.

Year:  2012        PMID: 23278024     DOI: 10.1063/1.4769753

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  Capsule implosions for continuum x-ray backlighting of opacity samples at the National Ignition Facility.

Authors:  Y P Opachich; R F Heeter; M A Barrios; E M Garcia; R S Craxton; J A King; D A Liedahl; P W McKenty; M B Schneider; M J May; R Zhang; P W Ross; J L Kline; A S Moore; J L Weaver; K A Flippo; T S Perry
Journal:  Phys Plasmas       Date:  2017-06-08       Impact factor: 2.023