Change of coronary flow by continuous-flow left ventricular assist device with cardiac beat synchronizing system (native heart load control system) in acute ischemic heart failure model.

Akihide Umeki1, Takashi Nishimura, Masahiko Ando, Yoshiaki Takewa, Kenji Yamazaki, Shunei Kyo, Minoru Ono, Tomonori Tsukiya, Toshihide Mizuno, Yoshiyuki Taenaka, Eisuke Tatsumi.   

Abstract

BACKGROUND: A novel control system for the EVAHEART left ventricular assist device (LVAD), known as the Native Heart Load Control System (NHLCS), can change the device's rotational speed (RS) in synchrony with the heartbeat. The system enhanced coronary flow (CoF) with the counter-pulse mode in normal goats' hearts, so we examined the change in CoF in goats with acute ischemic heart failure (HF). METHODS AND
RESULTS: We studied 14 goats (56.1±6.9kg) with acute ischemic HF induced by coronary microsphere embolization. We installed EVAHEART and ran the device in 4 modes [continuous support, circuit-clamp, counter-pulse (raise RS in diastole), and co-pulse (raise RS in systole)] with 50% or 100% bypass in each mode. In comparison with the circuit-clamp mode, CoF was 121.0±14.1% in the counter-pulse mode and 102.9±7.9% in the co-pulse mode, whereas it was 113.5±10.6% in the continuous mode, with 100% bypass (P<0.05). The same difference was confirmed with 50% bypass. The results indicated that a LVAD in an acute ischemic heart enhanced CoF, and that CoF was greater in the counter-pulse mode and smaller in the co-pulse mode relative to the continuous mode.
CONCLUSIONS: By using NHLCS to change CoF, recovery of native heart function with a LVAD has a better prognosis.

Entities:  

Mesh:

Year:  2012        PMID: 23269084     DOI: 10.1253/circj.cj-12-0676

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  3 in total

1.  Influence of a novel electrocardiogram-synchronized rotational-speed-change system of an implantable continuous-flow left ventricular assist device (EVAHEART) on hemolytic performance.

Authors:  Satoru Kishimoto; Kazuma Date; Mamoru Arakawa; Yoshiaki Takewa; Takashi Nishimura; Tomonori Tsukiya; Toshihide Mizuno; Nobumasa Katagiri; Yukihide Kakuta; Daisuke Ogawa; Motonobu Nishimura; Eisuke Tatsumi
Journal:  J Artif Organs       Date:  2014-09-02       Impact factor: 1.731

2.  Changing pulsatility by delaying the rotational speed phasing of a rotary left ventricular assist device.

Authors:  Kazuma Date; Takashi Nishimura; Mamoru Arakawa; Yoshiaki Takewa; Satoru Kishimoto; Akihide Umeki; Masahiko Ando; Toshihide Mizuno; Tomonori Tsukiya; Minoru Ono; Eisuke Tatsumi
Journal:  J Artif Organs       Date:  2016-07-19       Impact factor: 1.731

3.  Shifting the pulsatility by increasing the change in rotational speed for a rotary LVAD using a native heart load control system.

Authors:  Kazuma Date; Takashi Nishimura; Yoshiaki Takewa; Satoru Kishimoto; Mamoru Arakawa; Akihide Umeki; Masahiko Ando; Toshihide Mizuno; Tomonori Tsukiya; Minoru Ono; Eisuke Tatsumi
Journal:  J Artif Organs       Date:  2016-05-14       Impact factor: 1.731

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.