Lincomycin and spectinomycin combination therapy is widely used in veterinary medicine for the treatment of gastrointestinal and respiratory infections caused by lincomycin- and spectinomycin-sensitive microorganisms. A simple, reverse phase HPLC method for the analysis of samples of an injectable lincomycin and spectinomycin preparation containing a mixture of inactive excipients has been developed. The HPLC was carried out using the RP-C(18) (250 mm × 4.0 mm, 5 μm) column, with the gradient mobile phase consisting of an acetonitrile and phosphate buffer at pH 6; the flow rate of 1 mL/min and ultraviolet detection at 220 nm. This method was validated in accordance with both FDA and ICH guidelines and showed good linearity, accuracy, precision, selectivity, and system suitability results within the acceptance criteria. A stability-indicating study was also carried out and indicated that this method can also be used for purity and degradation evaluation of these formulations.Entities:
Keywords: Lincomycin; RP-HPLC; Spectinomycin; Stability-Indicating
Year: 2012 PMID: 23264944 PMCID: PMC3528054 DOI: 10.3797/scipharm.1207-13
Source DB: PubMed Journal: Sci Pharm ISSN: 0036-8709
Gradient elution programme
| 0–2 | 89 | 89 | 11 | 11 |
| 2–3 | 89 | 78 | 11 | 22 |
| 3–8 | 78 | 78 | 22 | 22 |
| 8–9 | 78 | 89 | 22 | 11 |
| 9–15 | 89 | 89 | 11 | 11 |
HPLC chromatographic conditions
| Flow rate | 1.0 mL/min |
| Wavelength (λ) | 220 nm |
| Stationary phase | RP18e, 5 μm, 250 × 4 mm |
| Column Temperature | 35°C |
| Injection volume | 20 μL |
| Run time | 15 minutes. |
Stress conditions
| Acidic hydrolysis | 1 mg/mL in 0.1N (up to 1N), HCl at 65 °C | 1–7 days |
| Basic hydrolysis | 1 mg/mL in 0.1N (up to 1N), NaOH at 65 °C | 1–7 days |
| Oxidative | 0.3% (up to 3%) H2O2; at RT; protected from light | Few hours to 7 days |
| Thermal | 70°C (RH of 40%) | Up to 3 weeks |
| Photolytic degradation | UV light (254 nm at room temperature) | Few hours to 3 days |
The applied ruggedness/robustness conditions
| pH values of the mobile phase | pH of the mobile 5.9, 6.0 & 6.1 |
| Detection wavelength | WL of 218, 220 and 222 nm |
| Flow rate of the mobile phase | Flow rate of 0.8, 1.0 and 1.2 mL/min |
| Elapsed assay times | The same analyst analysed the same trial in two different days |
| Analysts | Two analysts analysed the same trial in the same day |
Fig. 1.Linearity and range of spectinomycin sulphate
Fig. 2.Linearity and range of lincomycin hydrochloride
The results of specificity and stability-indicating studies.
| Acidic hydrolysis | Both major peaks area loss and degradation products (A) and (B). |
| Basic hydrolysis | Degradation product (C). |
| Oxidative | No change |
| Thermal | Loss of spectinomycin peak area |
| Photolytic degradation | No change |
Fig. 3.Chromatogram of well separated peaks of the active ingredient, degradative peaks, and the injections formulation.
Note: The small peaks prior to the active ingredients are degradative peaks and the late eluting peak is the preservative benzyl alcohol.
Accuracy and precision results of spectinomycin
|
| |||||||
|---|---|---|---|---|---|---|---|
|
| |||||||
| 1 | 250.00 | 251.00 | 250.50 | 253.00 | 251.00 | 252.00 | 99.40 |
| 2 | 250.00 | 251.00 | 250.50 | 99.40 | |||
| 3 | 250.00 | 250.00 | 250.00 | 99.21 | |||
|
| |||||||
|
| |||||||
| 1 | 311.00 | 311.00 | 311.00 | 312.00 | 312.00 | 312.00 | 99.68 |
| 2 | 315.00 | 315.00 | 315.00 | 100.96 | |||
| 3 | 314.00 | 313.00 | 313.50 | 100.48 | |||
|
| |||||||
|
| |||||||
| 1 | 376.00 | 376.00 | 376.00 | 375.00 | 375.00 | 375.00 | 100.27 |
| 2 | 376.00 | 376.00 | 376.00 | 100.27 | |||
| 3 | 374.00 | 374.00 | 374.00 | 99.73 | |||
| 99.93 | |||||||
| 0.59 | |||||||
| 0.59 | |||||||
Accuracy and precision results of Lincomycin
|
| |||||||
|---|---|---|---|---|---|---|---|
|
| |||||||
| 1 | 1204.00 | 1204.00 | 1204.00 | 1195.00 | 1193.00 | 1194.00 | 100.84 |
| 2 | 1201.00 | 1201.00 | 1201.00 | 100.59 | |||
| 3 | 1201.00 | 1201.00 | 1201.00 | 100.59 | |||
|
| |||||||
|
| |||||||
| 1 | 1485.00 | 1485.00 | 1485.00 | 1476.00 | 1477.00 | 1476.50 | 100.58 |
| 2 | 1497.00 | 1497.00 | 1497.00 | 101.39 | |||
| 3 | 1494.00 | 1493.00 | 1493.50 | 101.15 | |||
|
| |||||||
|
| |||||||
| 1 | 1770.00 | 1770.00 | 1770.00 | 1754.00 | 1754.00 | 1754.00 | 100.91 |
| 2 | 1772.00 | 1771.00 | 1771.50 | 101.00 | |||
| 3 | 1783.00 | 1783.00 | 1783.00 | 101.65 | |||
| 100.97 | |||||||
| 0.38 | |||||||
| 0.37 | |||||||
System suitability
| Resolution (R) | 2.4 | 6.7 | > 1.5 |
| Tailing (T) | 1.16 | 0.65 | ≤2.0 |
| Theoretical plates (N) | 3450 | 16500 | ≥1000 plates |