Literature DB >> 23264547

Patient safety in vitreoretinal surgery: quality improvements following a patient safety reporting system.

S Chien Wong1, Simon P Kelly, Paul M Sullivan.   

Abstract

AIMS: Vitreoretinal (VR) surgery is complex. Most clinical conditions that VR surgeons manage have a high risk for blindness or severe visual impairment. Reporting of patient safety incidents (PSI) in VR surgery was introduced at the Moorfields Eye Hospital (MEH) in the 1990s. We examine the role of PSI reporting in making VR surgery safer at our institution.
METHODS: Qualitative review of PSIs from 1997 to 2009 at MEH, London, UK.
RESULTS: Over the 13-year study period, 38 789 VR procedures were undertaken and 579 VR PSIs occurring in theatre or inpatient were reported. Mean rate of PSI reporting was 1.49% (range 0.12-3.35). In comparison, the mean rate of PSI reporting over the same period across all National Health Service ophthalmology in England was 0.59% (range 0.36-0.49). Overall, 0.9% of VR PSI resulted in 'Severe' harm, 11.6% in 'Moderate' harm and 87.5% in 'No' or 'Low' harm. 15 (2.6%) of PSIs directly resulted in a change in clinical practice, 13 of which occurred in the first half of the study period. 12 (3.6%) critical incidents were violations of pre-existing protocols and guidelines, eight of which occurred in the second half of the study period. 61 (10.9%) of PSIs fell into nine main error subtypes that contributed to a change in practice. The most common were ocular hypotony (2.9%), medical device failure (2.8%) and delay in VR surgery (1.2%).
CONCLUSIONS: VR PSI reporting resulted in a change in clinical practice. Longitudinal analysis suggests an accompanying increase in patient safety.

Entities:  

Mesh:

Year:  2012        PMID: 23264547     DOI: 10.1136/bjophthalmol-2012-301988

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  3 in total

Review 1.  How Effective Are Incident-Reporting Systems for Improving Patient Safety? A Systematic Literature Review.

Authors:  Charitini Stavropoulou; Carole Doherty; Paul Tosey
Journal:  Milbank Q       Date:  2015-12       Impact factor: 4.911

Review 2.  Factors that influence the recognition, reporting and resolution of incidents related to medical devices and other healthcare technologies: a systematic review.

Authors:  Julie Polisena; Anna Gagliardi; David Urbach; Tammy Clifford; Michelle Fiander
Journal:  Syst Rev       Date:  2015-03-29

3.  Case Series of Inappropriate Concentration of Intraocular Sulfur Hexafluoride.

Authors:  Piotr Kanclerz; Andrzej Grzybowski
Journal:  Case Rep Ophthalmol       Date:  2018-08-31
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.