Literature DB >> 23264012

[Comparison between the Airtraq, X-Lite, and direct laryngoscopes for thyroid surgery: a randomized clinical trial].

Mustapha Bensghir1, Chakib Chouikh, Najib Bouhabba, Salaheddine Fjjouji, Jalal Kasouati, Hicham Azendour, Noredine Drissi Kamili.   

Abstract

PURPOSE: To compare the performance of the Airtraq, X-Lite, and direct laryngoscopes for tracheal intubation for elective thyroid surgery.
METHODS: In this randomized prospective study, Patients who were ASA (American Society of Anesthesiologists) I or II physical status I or II and patients undergoing thyroid surgery were included in this randomized prospective study. Patients with attributes for difficult intubation criteria were excluded. After induction of anesthesia with fentanyl, propofol, and cisatracurium, patients were randomized into three groups according to the intubating device used: Airtraq laryngoscope (Airtraq), X-Lite video laryngoscope (X-lLite), and direct laryngoscope with a Macintosh blade (Direct). The primary outcome measurement was the time needed for tracheal intubation. Secondary outcomes included intubation difficulty score (IDS), Cormack-Lehane classification, as well as hemodynamic and respiratory effects.
RESULTS: One hundred and five patients were included. Mean (standard deviation) time for intubation was 28.2 (2.1) sec in the Airtraq group vs 36.6 (3.7) sec in the X-Lite group (P < 0.001), and 41.1 (4.4) sec in the Direct group (P < 0.001). Compared with to the direct laryngoscope, the Airtraq provided better visualization of the glottis (P = 0.003) with fewer attempts (P < 0.001) and lower IDS (P < 0.001). The intubation difficulty score was lower with the Airtraq than with the X-Lite (P = 0.003). Patients in the X-Lite and Direct groups showed more hemodynamic variations than those in the Airtraq group. There was only one failed intubation, in the direct group.
CONCLUSION: In thyroid surgery, the Airtraq and X-Lite laryngoscopes decrease time to intubation when compared with direct laryngoscopy.

Entities:  

Mesh:

Year:  2012        PMID: 23264012     DOI: 10.1007/s12630-012-9870-x

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  4 in total

Review 1.  Videolaryngoscopy versus direct laryngoscopy for adults undergoing tracheal intubation.

Authors:  Jan Hansel; Andrew M Rogers; Sharon R Lewis; Tim M Cook; Andrew F Smith
Journal:  Cochrane Database Syst Rev       Date:  2022-04-04

Review 2.  Videolaryngoscopy versus direct laryngoscopy for adult patients requiring tracheal intubation.

Authors:  Sharon R Lewis; Andrew R Butler; Joshua Parker; Tim M Cook; Andrew F Smith
Journal:  Cochrane Database Syst Rev       Date:  2016-11-15

3.  Case report of the use of videolaryngoscopy in thyroid goiter masses: An airway challenge.

Authors:  Stacey Watt; Jonathan Kalpan; Venkatesware Kolli
Journal:  Int J Surg Case Rep       Date:  2016-08-05

Review 4.  Video screen visualization patterns when using a video laryngoscope for tracheal intubation: A systematic review.

Authors:  Preston Dean; Benjamin Kerrey
Journal:  J Am Coll Emerg Physicians Open       Date:  2022-01-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.