Identification of conductive hearing loss in young infants using tympanometry and wideband reflectance.

Beth A Prieve1, Kathy R Vander Werff, Jonathan L Preston, Lea Georgantas.   

Abstract

OBJECTIVE: The goal of the study was to evaluate the effectiveness of tympanometry and wideband reflectance (WBR) in detecting conductive hearing loss (CHL) in young infants.
METHODS: Type of hearing loss was determined using auditory brainstem response using air- and bone-conducted tone bursts in 84 ears from 70 infants (median age = 10 weeks). Of these 84 ears, 60 are included in the current analysis: 43 with normal hearing (NH) and 17 with CHL. Tympanometry was measured using probe tone frequencies of 226, 678, and 1000 Hz. Tympanograms were evaluated in two ways: (1) Acoustic middle ear admittance (Ya, in millimhos); and (2) two-category classification (normal/abnormal), as described by Baldwin (2006). Measures of Ya were evaluated in two ways: by admittance-magnitude tympanograms and calculated admittance magnitude from subcomponents (conductance and susceptance). WBR was measured in response to a chirp stimulus after probe calibration. WBR was analyzed into thirteen 1/3 octave bands. Tests for statistical differences for two-category classification were analyzed using Chi-squared and Ya, and WBR were analyzed using repeated-measures analyses of variances. Cohen's d and likelihood ratios were computed for comparison with statistically significant differences.
RESULTS: Ya measured with 678- and 1000 Hz probe tones was significantly different between ears with CHL and NH. Two-category classification of tympanograms using a 1000 Hz probe tone was significantly different between ears with CHL and NH. Neither two-category classification nor Ya was significantly different between ears identified with CHL and NH using a 226 Hz probe tone. WBR was significantly higher in the frequency bands 800 to 2500 Hz and in the frequency band centered at 6300 Hz in infants with CHL. Effect sizes (Cohen's d) were greater than 2 for several WBR frequency bands and Ya measured with 1000 Hz probe tones. The results were similar for calculations of Ya from admittance-magnitude and subcomponent tympanograms. Positive likelihood ratios for WBR ranged between 8.1 and 38, and those for Ya using 1000 Hz ranged between 12.5 and 32.
CONCLUSIONS: CHL in young infants can be detected well with WBR or tympanometry using probe frequencies of 678 and 1000 Hz.

Entities:  

Mesh:

Year:  2013        PMID: 23263407     DOI: 10.1097/AUD.0b013e31826fe611

Source DB:  PubMed          Journal:  Ear Hear        ISSN: 0196-0202            Impact factor:   3.570


  5 in total

1.  Wideband absorbance tympanometry: a novel method in identifying otosclerosis.

Authors:  Arunraj Karuppannan; Animesh Barman
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-01-03       Impact factor: 2.503

2.  Acoustic Immittance, Absorbance, and Reflectance in the Human Ear Canal.

Authors:  John J Rosowski; Laura Ann Wilber
Journal:  Semin Hear       Date:  2015-02

3.  Longitudinal Development of Distortion Product Otoacoustic Emissions in Infants With Normal Hearing.

Authors:  Lisa L Hunter; Chelsea M Blankenship; Douglas H Keefe; M Patrick Feeney; David K Brown; Annie McCune; Denis F Fitzpatrick; Li Lin
Journal:  Ear Hear       Date:  2018 Sep/Oct       Impact factor: 3.570

4.  Rapid imaging of tympanic membrane vibrations in humans.

Authors:  Matan Hamra; Shadi Shinnawi; Mauricio Cohen Vaizer; Dvir Yelin
Journal:  Biomed Opt Express       Date:  2020-10-19       Impact factor: 3.732

5.  Longitudinal development of wideband reflectance tympanometry in normal and at-risk infants.

Authors:  Lisa L Hunter; Douglas H Keefe; M Patrick Feeney; Denis F Fitzpatrick; Li Lin
Journal:  Hear Res       Date:  2015-12-19       Impact factor: 3.208

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.