| Literature DB >> 23257942 |
Chiu Yuen To1, Yaser Badr, Boyd Richards.
Abstract
During treatment of a right internal carotid artery terminus aneurysm, an acute iatrogenic flow limiting dissection was caused in the cervical internal carotid. The true lumen was catheterized using a Mirage 0.008 microwire over an Excelsior SL-10 microcatheter, which was exchanged for a Marksman microcatheter. A 6 mm×30 mm Solitaire FR revascularization device was then deployed across the dissection as a salvage technique.Entities:
Mesh:
Year: 2012 PMID: 23257942 PMCID: PMC4545016 DOI: 10.1136/bcr-2012-010519