Phillip Bedggood1, Andrew Metha. 1. Department of Optometry and Vision Sciences, The University of Melbourne, 3010, Australia.
Abstract
Imaging the retinal vasculature offers a surrogate view of systemic vascular health, allowing noninvasive and longitudinal assessment of vascular pathology. The earliest anomalies in vascular disease arise in the microvasculature, however current imaging methods lack the spatiotemporal resolution to track blood flow at the capillary level. We report here on novel imaging technology that allows direct, noninvasive optical imaging of erythrocyte flow in human retinal capillaries. This was made possible using adaptive optics for high spatial resolution (1.5 μm), sCMOS camera technology for high temporal resolution (460 fps), and tunable wavebands from a broadband laser for maximal erythrocyte contrast. Particle image velocimetry on our data sequences was used to quantify flow. We observed marked spatiotemporal variability in velocity, which ranged from 0.3 to 3.3 mm/s, and changed by up to a factor of 4 in a given capillary during the 130 ms imaging period. Both mean and standard deviation across the imaged capillary network varied markedly with time, yet their ratio remained a relatively constant parameter (0.50 ± 0.056). Our observations concur with previous work using less direct methods, validating this as an investigative tool for the study of microvascular disease in humans.
Imaging the retinal vasculature offers a surrogate view of systemic vascular health, allowing noninvasive and longitudinal assessment of vascular pathology. The earliest anomalies in vascular disease arise in the microvasculature, however current imaging methods lack the spatiotemporal resolution to track blood flow at the capillary level. We report here on novel imaging technology that allows direct, noninvasive optical imaging of erythrocyte flow in humanretinal capillaries. This was made possible using adaptive optics for high spatial resolution (1.5 μm), sCMOS camera technology for high temporal resolution (460 fps), and tunable wavebands from a broadband laser for maximal erythrocyte contrast. Particle image velocimetry on our data sequences was used to quantify flow. We observed marked spatiotemporal variability in velocity, which ranged from 0.3 to 3.3 mm/s, and changed by up to a factor of 4 in a given capillary during the 130 ms imaging period. Both mean and standard deviation across the imaged capillary network varied markedly with time, yet their ratio remained a relatively constant parameter (0.50 ± 0.056). Our observations concur with previous work using less direct methods, validating this as an investigative tool for the study of microvascular disease in humans.
Entities:
Keywords:
(110.1080) Active or adaptive optics; (120.7250) Velocimetry; (170.1470) Blood or tissue constituent monitoring; (170.2655) Functional monitoring and imaging; (170.4470) Ophthalmology
The eye is the only organ in the body where the deep internal vasculature can be observed
directly, noninvasively and in high resolution. Correlation between abnormalities in the retinal
vasculature and abnormalities elsewhere in the body have been demonstrated in a diverse array of
conditions including diabetes [1], hypertension [2], stroke [3,4], Alzheimer’s disease [3,5], migraine [6] and glaucoma [7]. Each of these conditions
manifests pathology in the microvasculature [8-13], which is the primary site of exchange between blood and
tissue. Measurement of retinal blood flow, especially at the level of the microvasculature,
therefore offers a unique advantage for the longitudinal study of vascular disease in humans.Several techniques have become well established for the noninvasive measurement of blood flow in
the human retina. These include laser Doppler flowmetry [14],
laser speckle contrast imaging [15] and Doppler OCT [16]. These methods all rely on the relationship between the
velocity of red blood cells and the changes in frequency imparted to light that is scattered from
them [17]. These techniques provide repeatable measures of
blood flow in the larger retinal vessels, but suffer from limited spatial and temporal resolution as
well as low signal to noise ratio for vessels oriented more perpendicularly to the probe beam [18-20]. Erythrocyte
scatter also obfuscates the relationship between the frequency and velocity spectrum, limiting the
ability to measure absolute velocities via such indirect approaches [14]. These issues render the above techniques ineffective for study of the microvasculature,
although some approximations have been made [18,21].The required spatial resolution to image the microvasculature can be achieved using adaptive
optics, a technology that reshapes the optical wavefronts emanating from the eye [22] to achieve diffraction-limited resolution (<2 μm).
This is more than sufficient for visualization of the smallest retinal capillaries (~5 μm
diameter). Previous work using adaptive optics scanning laser ophthalmoscopy (AOSLO) has taken
advantage of the rarity of leukocytes in the blood stream to track them individually and so quantify
flow [23]. This approach revealed marked variability in
leukocyte velocity and pulsatility, between capillaries and between subjects, in normal retina
[24]. The most direct, or thoroughfare, capillaries were seen
to account for ~2/3 of all leukocyte traffic [25].
Thoroughfare capillaries were shown to preferentially undergo morphological changes in pre-clinical
diabetic retina, leading to leukocytes being redistributed to exchange capillaries [26] and confirming recent evidence of leukocyte-based microvascular
disruption in diabetes [27,28].Although useful to study in its own right, leukocyte flow is not a good surrogate for erythrocyte
flow due to the former’s larger size, reduced deformability [29] and relative rarity in the blood [30].
Erythrocyte flow has its own obvious importance in disease processes as the carrier of oxygen to
tissue. Erythrocyte flow in capillaries was not able to be measured unambiguously in previous
adaptive optics work due to limitations in spatial and temporal resolution [25]. An alternate approach in which the scanning raster of an AOSLO is fixed in one
dimension has allowed precise measurements to be made along the cross-section of larger vessels
~30-100 μm in diameter [31]. This has allowed the flow
profile to be characterized in these vessels, as well as changes to be measured resulting from the
cardiac cycle [32] and from visual stimulation of the retina
[33]. However this technique is also not suited to study of
erythrocyte flow in the microvasculature.To directly visualize erythrocytes in retinal capillaries we combined the spatial resolution
afforded by adaptive optics, the temporal resolution of a fast sCMOS camera, and a supercontinuum
laser that allows us to selectively deliver imaging light that is strongly absorbed by hemoglobin.
While confocal AOSLO systems typically focus on an intermediate position between the capillary bed
and the photoreceptor mosaic to aid image registration, our flood-based ophthalmoscope is able to
accurately register data while focused directly on the capillary bed. This overall approach allows
us to directly and unambiguously visualize the flow of individual erythrocyte packets, separated by
plasma gaps, in the retinal capillaries. This direct visualization of spatiotemporally dense and
high contrast moving particles lends itself to particle image velocimetry (PIV), which was used to
quantify erythrocyte flow.We provide details here on our methodology, the adaptation of standard PIV approaches to suit our
data, and validation of measured velocities against a noisy phantom data set. We use our approach to
report on the statistics of normal erythrocyte velocity and velocity fluctuations both within and
across capillaries in a normal human subject.
2. Methods
The subject, who is also the first author, was a healthy, nonsmoking, 29 year old male with no
ocular pathology or significant refractive error. The project was carried out in accordance with the
tenets of the Declaration of Helsinki, and approved by the University of Melbourne Human Ethics
Committee.
2.1. Adaptive optics imaging
The imaging system is a flood adaptive optics ophthalmoscope that has been described elsewhere
[34], with the exception that the retinal camera is now an
Andor NEO sCMOS device (Andor Technology PLC, Belfast, UK). The main functional difference brought
about by this change is the much greater frame rate. It is also somewhat more sensitive, with peak
sensitivity at 600 nm (0.57 quantal efficiency). The device operates in two distinct shutter modes:
rolling and global. Rolling shutter produces a staggered exposure between adjacent pixel rows (10
μs delay between each row) which allows the frame rate to be doubled in most cases. This
staggered exposure was deemed unsuitable for monitoring of blood flow in 2 dimensions due to the
high flow velocity, and we instead used global shutter which exposes all pixels simultaneously. The
device is capable of a sustained 50 fps at full frame (2560 x 2160) using global shutter, however
the frame rate can be greatly increased by restricting the region of interest in the vertical
direction, due to the way in which data is read out from the sensor. We limited the vertical extent
to 200 pixels (70 μm) to achieve a frame rate of 460 fps for this work. It was not necessary
to update the adaptive optics correction during the acquisition of our sequences, which were 60
frames (130 ms) in length.The sCMOS camera outputs a digital signal indicating the beginning of each frame exposure. This
output was used to trigger our imaging source, producing a train of exposures each 1.07 ms long. As
previously described [34], our imaging source is a
supercontinuum laser that is passed through acousto-optic tunable filters to select narrow wavebands
of interest for imaging, and then through 32 m of multimode fiber to minimize speckle. In this work
we selected wavebands centered on 593 nm. This provided near-optimal erythrocyte contrast, taking
into account the variance with wavelength of both the maximum permissible exposure (MPE) [35] and the output of our laser. Power at the cornea was 0.5 mW,
which is ~35 times below the MPE for each 130 ms imaging sequence according to ANSI guidelines
[35]. 5 imaging sequences were acquired in the same area and
presented for analysis in this work.Functional imaging of the cortex has shown changes in capillary flow to occur as quickly as 400
ms after the onset of neuronal activity [36]. Our 130 ms
acquisition time is therefore anticipated to be too short for a metabolic effect to occur as a
result of neuronal stimulation by the imaging light. Changes in inner retinal flow are also known to
be greatly reduced at high flicker rates [37], such as that
of our imaging light.
2.2. Imaged area
Figure 1
shows a conventional retinal fundus photograph from our subject (color), registered with a
motion-contrast enhanced [38] montage of the parafoveal
capillary network obtained at 593 nm, using our previous camera at 12 fps [39]. Approximately 100 frames were acquired at each 1.1° diameter area to
create the montage. Arterioles (A) and venules (V) are indicated on the figure. These were
identified based on color and tracking of branches from the optic nerve head, and confirmed by
observing the direction of flow in our sequences. The highlighted box and corresponding inset
indicates the area that was analyzed in this work, which is close to the edge of the foveal
avascular zone. This provided a capillary bed that was predominantly mono-stratified (to within ~0.1
D, or 35 μm) according to subjective judgment of capillary focus with our system; stepping
the focus through the entire retinal thickness did not reveal any additional vessels beyond this
layer at this retinal location. The area chosen is also a highly chaotic zone in which 2 nearby
arterioles deliver supply from opposite directions; the area displayed many sharp branch points and
sudden changes in velocity. The direction of flow in each capillary segment (red arrows) was
determined both subjectively and using PIV analysis (below), which were in excellent agreement.
Fig. 1
Area of analysis. Top: Conventional fundus image (color) overlaid with an adaptive optics montage
of the capillary network surrounding the foveal avascular zone (grayscale). A, V denotes arteries
and veins. Bottom: Shows inset demarcated by yellow border in top. Arrows indicate direction of
erythrocyte flow. Lower case letters denote capillary segments discussed in the text below. This is
a region of high confluence, with 2 neighboring arterioles (A) delivering blood in opposing
directions.
Area of analysis. Top: Conventional fundus image (color) overlaid with an adaptive optics montage
of the capillary network surrounding the foveal avascular zone (grayscale). A, V denotes arteries
and veins. Bottom: Shows inset demarcated by yellow border in top. Arrows indicate direction of
erythrocyte flow. Lower case letters denote capillary segments discussed in the text below. This is
a region of high confluence, with 2 neighboring arterioles (A) delivering blood in opposing
directions.
2.3. Image registration
After acquisition of image sequences, background noise was subtracted from the raw images. These
were then divided by a flat-field image, which was generated by collecting several hundred frames of
data at the same wavelength, at a slower frame rate, in which the subject was instructed to move
fixation erratically.Image sequences were registered using cross-correlation with sub-pixel accuracy. Prior to
registration, images were filtered in Fourier space to reduce noise, using a bandpass filter that
masked spatial details <0.7 μm or >18 μm in size. This filter was tapered using
a raised-cosine function (peaking at 1 in the image center and trailing to 0 at the edge) to reduce
the introduction of artifacts in the spatial image. Correction for angular rotation was necessary in
our previous work at lower frame rate [34], but was not
necessary with the current frame rate and short overall imaging time. Note that filtering was used
only to determine correct image registration; final data sequences for PIV analysis were not
filtered in this way.The above procedure was sufficient to provide high quality registration with the focus at the
level of the inner retinal capillaries, i.e. with minimal reliance on the underlying cone
mosaic.
2.4. Enhancement of erythrocyte motion
The static contrast of capillaries and erythrocytes, relative to background tissue, was low. This
is illustrated in Fig. 2
(top), which shows a raw frame from one of our data sequences. PIV approaches typically make
use of a high contrast tracer particle [40] that can easily
be divorced from the surrounding media; an analogue for measurement of retinal flow is fluorescein
angiography [41], however this is an invasive technique.
Fortunately the moving erythrocytes in our sequences produced large changes in contrast over time;
this is best appreciated in the sequence Fig. 2
(Media 1), or by
the “division” image analysis [38] of the same
sequence in Fig. 2 (bottom). A useful, direction-preserving
method to accentuate such changes in the context of PIV was to subtract the first principal
component from each pixel, using time as a variable. This was found to greatly eliminate static
structural information while leaving erythrocyte motion intact; we filtered each data sequence in
this way before analysis.
Fig. 2
Sample of our imaging data. Top: a single frame at 593 nm, 1° nasal and 2° inferior
to the foveal center, close to the edge of the foveal avascular zone. Bottom: Motion contrast
enhancement of the same area, based on 60 frames of data (130 ms). The sequence itself is shown in
Media 1.
Sample of our imaging data. Top: a single frame at 593 nm, 1° nasal and 2° inferior
to the foveal center, close to the edge of the foveal avascular zone. Bottom: Motion contrast
enhancement of the same area, based on 60 frames of data (130 ms). The sequence itself is shown in
Media 1.Another possible approach to enhance erythrocyte contrast would be to use multiple wavelengths to
exploit the spectral absorption signature of hemoglobin, but we have previously found this to be of
dubious reliability at the capillary level when viewed in vivo [39]. This approach is presumably frustrated by complex interference
and scattering effects which outweigh wavelength-selective absorption in single erythrocytes, which
are individually of low optical density.
2.5. Particle image velocimetry
In PIV, each frame of the sequence is divided into small regions of interest (ROIs) and
cross-correlated with a larger surrounding ROI from the previous frame [40]. The displacement of the peak of the cross-correlation function indicates the
ensemble motion of particles present within the region. Larger ROIs correspond to faster maximum
measurable velocity, while smaller ROIs correspond to better spatial resolution; the latter is
important for measuring velocity changes along a vessel, while avoiding the influence of nearby
vessels and crossing points. It is therefore important to select the ROI size based on known
physiological constraints [40]. It is possible to expand the
range of measurable velocities by employing search boxes of varying size, however this requires that
a method be devised to select the most appropriate measurement at each position in the image.It is apparent that there is no fixed ideal set of parameters for PIV analysis; the parameters
reported here were chosen using trial and error to empirically maximize accuracy when applied to a
known, noisy (“phantom”) data set—see section 2.6 below. Each step in the
algorithm was found to be useful in defeating a particular sub-class of errors in the phantom data
set. The maximum velocity measurable was beyond the range of physiologically relevant capillary
velocities [24,25,42], to ensure that no data was lost as a result
of insufficient ROI size. The final parameters of our algorithm were as follows:Minimum velocity: 1.5 pixels/frame (0.24 mm/s)Maximum velocity: 15, 25, or 30 pixels/frame (2.4, 4.0, 4.8 mm/s)—3 separate analysesSmall ROI size: 32 pixels (11. 2 x 11.2 μm)Large ROI size: 48, 80 or 96 pixels (17, 28 or 34 μm)—chosen based on max.
velocitySince the axis of motion at each point on a capillary cannot change, we arrived at an overall
direction of motion for each sample point by taking the median flow angle over an image sequence.
Vectors more than 30° from this value were rejected as noise-evoked outliers, and the median
angle of the remaining vectors taken again. To determine speed, these vectors were projected onto
the direction found above, and the median of the resulting lengths was taken. A minimum of 6
measurements were required for a point to be considered valid data, to minimize background noise.
The final array of flow vectors was passed through a 2D median filter, which further suppressed
background noise. This stringent approach had the disadvantage that no data was recorded for some
capillary segments.To determine how velocity varied over shorter time scales we employed a rolling temporal window
of 15 frames (33 ms) duration, which is equivalent to passing the velocity signal through a lowpass
filter. The same procedure as above was applied, however the initial direction of motion as
determined for all frames was used as the starting direction. This greatly assisted in correctly
rejecting spurious motion in this much smaller number of measurements, however, it did leave the
algorithm blind to any rapid reversals in motion. Visual inspection of our data did not reveal any
reversals to occur, and if they did, these would be visible as gaps in the velocity trace for a
given capillary segment.Analysis with three different maximum velocities was used to increase the range of valid
measurable velocities. A robust method to select the most appropriate velocity, determined
empirically (see below), was to simply accept the fastest of the 3 measurements.
2.6. Phantom data set
We generated a model, or phantom, data set to allow us to empirically determine a robust approach
for velocity estimation from quantifiably noisy data. We used software to construct an imaging field
of the same nominal size, consisting of straight, invisible “vessels” oriented in
various directions on a white background. Each vessel was populated with black rectangular
“erythrocyte” packets (5.0 x 12 μm in dimension), with a 12 μm gap
between each packet. In succeeding frames of the phantom sequence these packets were propagated at
constant speed along each vessel, with the speed for different vessels chosen to span the known
physiological range of capillary velocities [24,25,42]. To each pixel of
each binary frame was added a sufficient level of white noise such that the PIV approach of section
2.5 rejected a similar percentage of frame pairs as for our actual image data. In this way we
regarded the phantom set as being sufficiently degraded in image quality so as to mimic the real
data.Figure 3
(grayscale) shows a single frame from a phantom data sequence, while the colored arrows show
the PIV analysis for the entire 60 frame sequence. Both arrow color and length indicate the measured
velocity. Points devoid of coloured arrows did not pass the acceptability criteria described in
section 2.5; i.e. the data was not differentiated from noise. The numbers listed over each vessel
segment show median velocity ± median absolute difference of the line of points closest to
the center of each vessel. The numbers in parantheses indicate the true (input) velocity.
Measurements close to a crossing point were ignored. There were a minimum of 10 readings for each
vessel. The error rate ranged from 0 to 7% from the true velocity, confirming the robustness of this
approach in the face of a noisy data set. We are therefore confident to state that, away from branch
and crossing points, our margin for error in measured velocity in our actual imaging sequences is
<10%.
Fig. 3
Phantom data set for validation of measured velocities, consisting of 7 modeled capillaries.
Grayscale: a single frame from the phantom data sequence. Modeled erythrocyte clusters are visible
as dark rectangular patches. Color overlay: Velocity vectors in each region of interest. Speed is
denoted both by arrow length and color. Numbers: median velocity ± median absolute difference
(true velocity). Units are mm/s. Velocity measurements were derived by considering a line of points
(minimum 10 per vessel) that were closest to the vessel center.
Phantom data set for validation of measured velocities, consisting of 7 modeled capillaries.
Grayscale: a single frame from the phantom data sequence. Modeled erythrocyte clusters are visible
as dark rectangular patches. Color overlay: Velocity vectors in each region of interest. Speed is
denoted both by arrow length and color. Numbers: median velocity ± median absolute difference
(true velocity). Units are mm/s. Velocity measurements were derived by considering a line of points
(minimum 10 per vessel) that were closest to the vessel center.
2.7. Reporting of velocity
A potential limitation of our measurements is that motion is only measured transversely in the
retina (i.e., along the surface of our detector) as opposed to along the true capillary detection,
which may contain an axial component. This is a minor limitation because most capillaries are
oriented almost entirely transversely in the retina [43], and
strong exceptions to this rule should be readily apparent from our sequences by changes in
erythrocyte packet length along a given capillary. It was not possible to detect any such changes in
the area that we imaged here.It should be noted that even though data is reported only for a single subject, the large number
of erythrocytes contributing to each velocity measurement in each capillary afford good reliability
compared with leukocyte tracking methods [24,25] which rely on a smaller number of samples due to the scarcity
of leukocytes in the blood stream.Measurements of particle displacement were converted into absolute distances by adjusting for
measured axial length of the subject’s eye, which was 23.97 mm. This was measured with an
Axis II ultrasound device (Quantel Medical, Clermont-Ferrand, France).
3. Results
Figure 4
shows the median velocity over the course of the same data sequence from Fig. 2 (Media 1).
Velocity is represented both by arrow length tangential to the vessel orientation, and by arrow
color. Spatial variability was high even over the small area shown (~0.8 x 0.2°), with
erythrocyte velocities ranging from ~0.5-3.1 mm/s across space in this sequence. The variability can
be further appreciated in Fig. 2
(Media 1).
Fig. 4
PIV analysis (median velocity) from the same image sequence as in Fig. 2 (Media 1).
Imaging wavelength was 593 nm, imaged area was 1° nasal and 2° inferior to the foveal
center. Diameter of displayed region is ~0.8 x 0.2°. Grayscale: motion contrast enhanced
“division” image [38] from this sequence. Color
arrows: Velocity vectors in each region of interest. Speed is denoted both by arrow length and
color. Letters: Labeled capillary segments referred to in the text; labels correspond to those in
Fig. 1.
PIV analysis (median velocity) from the same image sequence as in Fig. 2 (Media 1).
Imaging wavelength was 593 nm, imaged area was 1° nasal and 2° inferior to the foveal
center. Diameter of displayed region is ~0.8 x 0.2°. Grayscale: motion contrast enhanced
“division” image [38] from this sequence. Color
arrows: Velocity vectors in each region of interest. Speed is denoted both by arrow length and
color. Letters: Labeled capillary segments referred to in the text; labels correspond to those in
Fig. 1.It is instructive to divide the capillaries into segments; we defined a length of vessel as a
capillary segment if it consisted of at least 5 data points along the vessel that were unbroken by
missing data or a branch/crossing, and in which single file erythrocyte flow was subjectively
apparent. We noted cases of marked variability along several capillary segments, including between
consecutive segments where no branching had occurred. For example, in Fig. 4, segment u shows markedly reduced velocity compared to the
upstream and downstream segments j and a. This behavior is assumedly a
result of the crossing vessels bounding segment u. It is also interesting to note the
downstream increase in speed along segment a, which must indicate either widening of
the capillary segment, or erythrocytes “catching up” to the less impeded [44] flow of plasma. In a separate sequence (not shown), instead
segment u maintained a relatively constant velocity (1.55 ± 0.09 mm/s), while
the surrounding segments were far more variable (1.51 ± 0.48 and 1.23 ± 0.32 mm/s for
j and a). In other words, in that sequence segment u was
likely buffered against changes in velocity by traffic in the crossing segments.Figure 5
shows the standard deviation in velocity over the course of a sequence that showed high
temporal variability. Arrow length and color both denote the standard deviation in velocity, with
each individual time point incorporating 15 frames of data as explained above. The variability
inherent in this sequence can be further appreciated in Fig.
5 (Media 2), which
shows the original data sequence together with an overlay of the rolling temporal PIV analysis. Flow
in the entire capillary network in this area can be seen to increase dramatically
Fig. 5
PIV analysis (standard deviation) from an image sequence showing marked variability of flow over
time, as well as location. The corresponding sequence is shown in Media
2. Grayscale: motion contrast enhanced “division”
image [38] from this sequence. Color arrows: Standard
deviation of velocity in time, using temporal windows 15 frames (33 ms) long. Standard deviation is
denoted both by arrow length and color. Letters: Labeled capillary segments referred to in the text;
labels correspond to those in Fig. 1.
PIV analysis (standard deviation) from an image sequence showing marked variability of flow over
time, as well as location. The corresponding sequence is shown in Media
2. Grayscale: motion contrast enhanced “division”
image [38] from this sequence. Color arrows: Standard
deviation of velocity in time, using temporal windows 15 frames (33 ms) long. Standard deviation is
denoted both by arrow length and color. Letters: Labeled capillary segments referred to in the text;
labels correspond to those in Fig. 1.over the course of the sequence. To derive an overall velocity measure for each capillary
segment, we took the median of all data points that lay along the vessel and were oriented along the
apparent vessel direction on the motion contrast image. Figure
6
shows the result of this analysis, in which the velocity in each segment is seen to ramp up
throughout the duration of the sequence. It is worth noting that velocity in consecutive segments
u and a (plotted in red) increased dramatically, but with the change in
downstream segment a preceding upstream segment u by some 40 ms. This
presumably corresponds to upstream “traffic” in the venous direction that has caused a
back-propagation of resistance up the vessel.
Fig. 6
Velocity in each capillary segment in the same image sequence shown in Fig. 5 (Media 2).
Velocity generally increased throughout the course of this sequence. Blue plots indicate capillary
segments e, g, m and w that were present in all
sequences, and are analyzed in Fig. 7. Red plots indicate
segments a (asterisks) and u (crosses), which are consecutive and are
referred to in the text.
Velocity in each capillary segment in the same image sequence shown in Fig. 5 (Media 2).
Velocity generally increased throughout the course of this sequence. Blue plots indicate capillary
segments e, g, m and w that were present in all
sequences, and are analyzed in Fig. 7. Red plots indicate
segments a (asterisks) and u (crosses), which are consecutive and are
referred to in the text.
Fig. 7
Velocity in the same 4 capillary segments across 5 image sequences. Each panel corresponds to the
same capillary segment, and each plot color corresponds to the same image sequence.
The global correlation in velocity evident in Fig. 6 was
not apparent in the other sequences. To illustrate this, Fig.
7
shows the velocity plots for each of the 4 segments plotted in blue in Fig. 6. These segments (e, g, m and w) were
visible in all 5 sequences analyzed. Each panel of Fig. 7
corresponds to a segment, and each plot to a data sequence. Of interest is that a) a given capillary
segment was able to have a diverse array of mean velocity and variability in velocity; b) in the
same image sequence some capillaries changed velocity markedly while others remained highly
constant; and c) during the sequence of generally increasing velocity, the segments did not all
increase their velocity at the same time.Velocity in the same 4 capillary segments across 5 image sequences. Each panel corresponds to the
same capillary segment, and each plot color corresponds to the same image sequence.Although we have presented data here only from selected sequences that provide the most clear and
illustrative examples, interested readers may obtain the entirety of our velocity data upon
request.
4. Discussion
We have demonstrated and validated a technique that allows direct, noninvasive visualization of
erythrocyte packets flowing in single file through capillaries in the living human retina. The range
of erythrocyte velocities in the measured area (0.3–3.3 mm/s range, mean 1.3 mm/s) agrees
well with previously published figures for retinal leukocytes in humans [24,25] and for cerebral erythrocytes in
animals (see [42] for review). Our observation of marked
variability in the distribution of erythrocyte velocities across space and time also echoes
observations made by others [45-47].Additional observations of interest, noted above, included: the potential for global velocity in
the imaged area to correlate between all capillary segments, which was likely related to cardiac
activity; slowing and buffering of changes to flow resulting from contact by neighbouring
capillaries, which was likely due to compressive effects on the capillary lumen; variations in
velocity along capillary segments in the absence of any branching, due either to alterations in
resistance along short lengths of a capillary or to differences between plasma and erythrocyte
velocity; and changes in velocity propagating upstream in a vessel, likely due to downstream
“traffic”.Passive fluctuations in capillary flow are thought to occur as a result of interaction between
the vessels (geometry, redundancy and compliance) and the stochastic variations that occur in the
number and size of erythrocytes [44,46,48]. This causes large fluctuations to
occur at rates that are not readily comparable to the heart or respiration rates—full range
fluctuations have been noted to occur as quickly as 40-90 Hz [46], or as slow as 0.1 Hz [45]. In other words,
although heart and respiration rates are an important source of variability in capillary velocity
[25], it is by no means a straightforward relationship; large
amounts of data are potentially required to meaningfully characterize mean velocity or pulsatility
of each capillary. This somewhat limits the usefulness of analyzing the flow dynamics of any single
capillary.A more practical approach may be to analyze the overall network properties of the capillary
plexus in a given location. Mounting evidence suggests that the heterogeneity of flow in the
capillary network is adjusted based on the metabolic needs of the tissue, with flow being
redistributed to more homogeneous patterns under metabolic challenge [42,47,49-52]. Modeling predicts this redistribution
to significantly improve metabolite exchange with tissue, especially oxygen extraction [53]. Redistribution of flow may result from constriction/dilation
of pericytes [54] or pre-capillary arterioles [55]. If capillary heterogeneity is indeed linked to metabolic
needs, a useful measure of heterogeneity should vary little in a given retinal location under the
same adaptation and environmental conditions, as in our experiment. Comparing instantaneous velocity
for the imaged capillary network across all time points in our sequences, mean network velocity was
1.33 ± 0.28 mm/s, while standard deviation in network velocity was 0.65 ± 0.12 mm/s.
However the ratio of these parameters (standard deviation / mean) was more constant with time (0.50
± 0.056). This implies that under constant metabolic need, transient increases in network
velocity are accompanied by commensurate increases in the variability between capillary segments.
Further work is currently underway to establish the range of normative values for capillary
heterogeneity in healthy subjects both at baseline and under metabolic activity (e.g.
hypoxia/hypercapnia and functional stimulation). This will allow comparisons to be made to cases of
vascular disease, to learn how baseline and autoregulatory dynamics in flow may be compromised at
the capillary level.The rules governing division of flow at capillary branches are of interest. If it is assumed that
the local haematocrit does not change significantly after a branch, Kirchoff’s circuit law
suggests that the velocity prior to the branch should be the sum of the velocities after the branch.
This was observed to be the case in certain vessels; for example, segment e branches
into segments d and g, with the component velocities summing to within 1%
of e in one sequence and within 10% in another sequence. However this assumption was
violated in other segments; for example in one sequence the branches m and
n from segment k gave only 70% of the total velocity of segment
k. Therefore it is apparent that the local haematocrit must be allowed to vary between
branches, as has been reported by others [47,50,51]. This result is
expected from theory—in fact, in the absence of stochastic variations in erythrocyte size and
position within the capillary, all erythrocytes would follow the branch offering the lowest
resistance [44]. Variations in haematocrit mean that
measurement of velocity alone is insufficient to completely characterize the dynamics of capillary
flow. It should be possible for local haematocrit data to be extracted from our sequences, for
example by counting intensity minima passing each point along a capillary. There is still some
uncertainty in this approach, since the number of erythrocytes comprising each packet is not known
(see below). For this reason a more practical measure may be to simply quantify the proportion of a
capillary segment that is occupied by erythrocytes.We cannot state whether each and every moving erythrocyte packet in our sequences (e.g. see Fig. 2, Media
1) is a single erythrocyte, or a small number of erythrocytes
traveling in apposition. Erythrocyte diameter is ~8 μm when not being forced through the
capillary bed, while measured packet length was ~10-12 μm. Erythrocytes are known to lengthen
somewhat in the capillaries [56], as a result of their
relatively liquid center [48] and the ~5 μm capillary
diameter. It should also be noted that we expect some artifactual enhancement of apparent packet
length due to motion blur over our ~1 ms exposure time, on the order of ~0.5-3 μm according
to our measured range of erythrocyte speeds. Given these considerations, it would be instructive to
quantify erythrocyte packet length in different capillary segments. In practice to date, we have
found this difficult because a given packet often changes in appearance between high and low
contrast as it traverses the capillary bed. This could result from changes in specular reflection as
the deformable packet conforms to the shape of the capillary wall, or from interference between the
anterior and posterior faces of the capillary wall (which are only 5 μm apart and so within
the ~7 μm coherence length of our imaging light source [34]). Thus our rough estimate of packet length stems from a relatively small number of
packets, taken between adjacent frames where the packet appearance remained at high contrast. We
have therefore presented results here only for analysis of packet velocity, to which the PIV method
is well suited.A primary limitation to consider for this technique is the relatively short imaging wavelength
that was used to maximize erythrocyte contrast (593 nm), compared to the longer wavelengths that are
more typically used to enhance patient comfort and reduce the risk of light exposure damage. To
determine whether such a short imaging wavelength was truly necessary, we repeated our imaging
procedure using 720 nm light. Dark, moving erythrocyte packets were still apparent, with reduced
signal to noise ratio, such that our PIV algorithm obtained useful velocity measures in only
~1/2–2/3 of capillary segments compared with 593 nm. The 720 nm data was also much more
sensitive to small changes in image quality. Despite the reduced signal to noise ratio, contrast was
in fact much higher than expected based solely on the ratio of hemoglobin absorption between these
wavelengths, which is more than an order of magnitude. This suggests that scatter and/or
interference [39] effects dwarf the absorption of light due
to hemoglobin alone, when imaging individual erythrocyte packets in the capillaries. This limits the
measurement of hemoglobin content via spectral means in these smallest of vessels. For the purposes
of improving contrast for velocity measurement, any interference present may be enhanced by
increasing the spatial coherence (confining the illumination delivery within the pupil) and the
temporal coherence (reducing the bandwidth) of the illumination.It is worth noting that after the desired focus was set, all acquired sequences were used for
analysis. Thus the PIV method did not seem to have any stringent image quality requirement over and
above the typical AO imaging work that we have undertaken in the past [34,39]. In other words, efficiency of data
collection was high, which mitigates concerns about the somewhat lower MPE at 593 nm. To further
mitigate these concerns we allowed an average of 2-3 minutes to elapse between acquisition of each
sequence. To achieve more efficient data collection in a larger number of subjects, it may be
prudent to acquire sequences from each desired retinal eccentricity in series, repeating this entire
process as many times as desired, to further minimize exposure of each individual area of the
retina. To prevent loss of data quality it is recommended to restrict analysis to areas in which
there is effectively single stratification of the capillary bed, for example in the region
immediately surrounding the foveal avascular zone.An important step in the application of PIV to our data was the rejection of spurious individual
velocity measurements, based on presumed physiological constraints on erythrocyte speed and the
direction of flow. However under the influence of disease and age, there is potential for marked
slowing or acceleration of erythrocyte velocity and/or the formation of abnormal capillary geometry
(e.g. micro-aneurysms, anastomoses). This may necessitate that outliers be identified by statistical
methods [57-58], rather than preconceptions of the likely range of values. Beyond these limitations,
data acquisition will be limited in ways typical for flood adaptive optics imaging in general,
namely by the presence of scattering media and small pupils in the aged eye, poor fixation, or large
amounts of higher order aberrations that are not well corrected by the deformable mirror. Data from
more subjects are required to evaluate the influence of aged, diseased or otherwise difficult eyes
on the approach detailed here.The PIV and division [39] approaches discussed above
successfully identified the majority of capillary segments in our sequences. However the signal to
noise ratio was observed to be low in several capillary segments in which motion was nonetheless
subjectively identified with relative ease. This was especially true in regions that approached
branch points and neighboring vessels. While the reliability of such subjective analysis is
debatable, it is possible that approaches can be borrowed from investigations of complex human
motion perception and used to more accurately characterize the blood flow statistics of these
relatively low contrast data sets.
5. Conclusions
We have devised a method to directly and noninvasively visualize erythrocytes flowing in single
file in capillaries in the living human eye. The method lends itself well to particle image
velocimetry for the quantification of erythrocyte velocity, which has confirmed high spatiotemporal
variability in flow of erythrocytes in retinal capillaries. Our method complements other tools for
the assessment of blood flow in the retina, which include assessment of retinal leukocyte dynamics
[25,26], erythrocyte
flow profiles in arterioles and venules [31,32], and velocity in the larger retinal vessels via Doppler/speckle
measurement [14-16]. Together these tools allow the noninvasive study of flow dynamics in retinal vessels of
any caliber in the living human eye, which has important ramifications not only for study of retinal
disease but also for diseases of the cerebrovascular and cardiovascular systems.
Authors: Brian White; Mark Pierce; Nader Nassif; Barry Cense; B Park; Guillermo Tearney; Brett Bouma; Teresa Chen; Johannes de Boer Journal: Opt Express Date: 2003-12-15 Impact factor: 3.894
Authors: Stephen A Burns; Ann E Elsner; Toco Y Chui; Dean A Vannasdale; Christopher A Clark; Thomas J Gast; Victor E Malinovsky; Anh-Danh T Phan Journal: Biomed Opt Express Date: 2014-02-27 Impact factor: 3.732
Authors: Toco Y P Chui; Michael Dubow; Alexander Pinhas; Nishit Shah; Alexander Gan; Rishard Weitz; Yusufu N Sulai; Alfredo Dubra; Richard B Rosen Journal: Biomed Opt Express Date: 2014-03-13 Impact factor: 3.732
Authors: Alexander Pinhas; Moataz Razeen; Michael Dubow; Alexander Gan; Toco Y Chui; Nishit Shah; Mitul Mehta; Ronald C Gentile; Rishard Weitz; Joseph B Walsh; Yusufu N Sulai; Joseph Carroll; Alfredo Dubra; Richard B Rosen Journal: Invest Ophthalmol Vis Sci Date: 2014-11-20 Impact factor: 4.799
Authors: Boyu Gu; Xiaolin Wang; Michael D Twa; Johnny Tam; Christopher A Girkin; Yuhua Zhang Journal: Biomed Opt Express Date: 2018-07-12 Impact factor: 3.732
Authors: Woo June Choi; Wan Qin; Chieh-Li Chen; Jingang Wang; Qinqin Zhang; Xiaoqi Yang; Bruce Z Gao; Ruikang K Wang Journal: Biomed Opt Express Date: 2016-06-20 Impact factor: 3.732