Anterior segment optical coherence tomography versus ultrasound pachymetry to measure corneal thickness in endothelial keratoplasty donor corneas.

Ryan J Fante1, Roni M Shtein, Michael S Titus, Maria A Woodward.   

Abstract

PURPOSE: To analyze corneal thickness (CT) by anterior segment optical coherence tomography and ultrasound pachymetry in precut endothelial keratoplasty (EK) donor corneas at the eye bank.
METHODS: Thirty-six corneas were analyzed. All corneas were dissected to create lamellar grafts for EK using the standard eye bank protocol. We measured central and peripheral (3 mm from center) CT by ultrasound after lamellar tissue processing and by optical coherence tomography after the tissue was transferred to a viewing chamber. We performed paired t tests and Pearson correlation analyses to compare ultrasound and optical coherence tomography measurements.
RESULTS: Central CT measurements by optical coherence tomography versus ultrasound in donor corneas were not statistically different (mean: 169 vs. 177 µm, respectively, P = 0.09). There was a strong correlation between measurements of central CT (r = 0.73, P < 0.0001). There was a statistically significant difference in mean peripheral CT measurements by optical coherence tomography and ultrasound (mean: 190 vs. 236 µm, respectively, P < 0.0001); at peripheral locations, measurements correlated moderately (r = 0.52, P = 0.002).
CONCLUSIONS: Optical coherence tomography measurements of EK-prepared tissue averaged 8 μm thinner than the ultrasound measurements, similar to clinical measurements of central CT. Given the growing interest in the effect of EK graft thickness on patient outcomes, it is important to understand the differences in measurements obtained by different devices used to assess donor corneas.

Entities:  

Mesh:

Year:  2013        PMID: 23238397     DOI: 10.1097/ICO.0b013e31827c0b5b

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  2 in total

Review 1.  Eye-bank preparation of endothelial tissue.

Authors:  Grace E Boynton; Maria A Woodward
Journal:  Curr Opin Ophthalmol       Date:  2014-07       Impact factor: 3.761

2.  Accuracy of corneal trephination depth using the Moria single-use adjustable depth vacuum trephine system.

Authors:  Carlton R Fenzl; Adam J Gess; Majid Moshirfar
Journal:  Clin Ophthalmol       Date:  2014-11-27
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.