Literature DB >> 23234809

Dense arrays of micro-needles for recording and electrical stimulation of neural activity in acute brain slices.

D E Gunning1, J M Beggs, W Dabrowski, P Hottowy, C J Kenney, A Sher, A M Litke, K Mathieson.   

Abstract

OBJECTIVE: This paper describes the design, microfabrication, electrical characterization and biological evaluation of a high-density micro-needle array. The array records from and electrically stimulates individual neurons simultaneously in acute slices of brain tissue. APPROACH: Acute slices, arguably the closest in-vitro model of the brain, have a damaged surface layer. Since electrophysiological recording methods rely heavily on electrode-cell proximity, this layer significantly attenuates the signal amplitude making the use of traditional planar electrodes unsuitable. To penetrate into the tissue, bypassing the tissue surface, and to record and stimulate neural activity in the healthy interior volume of the slice, an array of 61 micro-needles was fabricated. MAIN
RESULTS: This device is shown to record extracellular action potentials from individual neurons in acute cortical slices with a signal to noise ratio of up to ∼15:1. Electrical stimulation of individual neurons is achieved with stimulation thresholds of 1.1-2.9 µA. SIGNIFICANCE: The novelty of this system is the combination of close needle spacing (60 µm), needle heights of up to 250 µm and small (5-10 µm diameter) electrodes allowing the recording of single unit activity. The array is coupled to a custom-designed readout system forming a powerful electrophysiological tool that permits two-way electrode-cell communication with populations of neurons in acute brain slices.

Mesh:

Year:  2012        PMID: 23234809     DOI: 10.1088/1741-2560/10/1/016007

Source DB:  PubMed          Journal:  J Neural Eng        ISSN: 1741-2552            Impact factor:   5.379


  2 in total

1.  A Materials Roadmap to Functional Neural Interface Design.

Authors:  Steven M Wellman; James R Eles; Kip A Ludwig; John P Seymour; Nicholas J Michelson; William E McFadden; Alberto L Vazquez; Takashi D Y Kozai
Journal:  Adv Funct Mater       Date:  2017-07-19       Impact factor: 18.808

Review 2.  Future of Smart Cardiovascular Implants.

Authors:  Anubhav Bussooa; Steven Neale; John R Mercer
Journal:  Sensors (Basel)       Date:  2018-06-22       Impact factor: 3.576

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.