Comparison of temporal artery, nasopharyngeal, and axillary temperature measurement during anesthesia in children.

Sevtap Hekimoglu Sahin1, Ridvan Duran, Necdet Sut, Alkin Colak, Betul Acunas, Burhan Aksu.   

Abstract

STUDY
OBJECTIVE: To evaluate the accuracy and precision of a new, noninvasive infrared thermometer applied to the temporal artery.
DESIGN: Prospective randomized study.
SETTING: Trakya University Hospital. PATIENTS: 60 ASA physical status 1 and 2 children undergoing surgery.
INTERVENTIONS: During anesthesia, temperature measurements were recorded with three different techniques: temporal artery, nasopharynx, and axillary temperature. MEASUREMENTS: Temperatures measured from the nasopharynx, temporal artery, and the axilla were recorded at 15-minute intervals for the first hour, then at 30-minute intervals until the completion of surgery. During each measurement, heart rate and midarterial pressure were recorded. MAIN
RESULTS: There were no statistically significant differences between temperatures recorded at the temporal artery and nasopharynx at 15, 30, 45, 60, 90, and 120 minutes, and the completion of surgery. Axillary temperatures were statistically lower than those recorded at the nasopharynx and the temporal artery (P < 0.001). Bland-Altman plots showed a correlation of temperature measurements between the temporal artery and nasopharyngeal methods. The axillary method had a lower correlation with the temporal artery and the nasopharyngeal methods.
CONCLUSIONS: The temporal artery thermometer is a substitute for the nasopharyngeal thermometer for core temperature measurement during anesthesia in children.
Copyright © 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23228869     DOI: 10.1016/j.jclinane.2012.05.003

Source DB:  PubMed          Journal:  J Clin Anesth        ISSN: 0952-8180            Impact factor:   9.452


  5 in total

1.  Examination of intra-operative core temperature in joint arthroplasty: a single-institution prospective observational study.

Authors:  Jennifer R Matos; Julie R McSwain; Bethany J Wolf; J Wesley Doty; Sylvia H Wilson
Journal:  Int Orthop       Date:  2018-05-11       Impact factor: 3.075

2.  Are Non-Contact Thermometers an Option in Anaesthesia? A Narrative Review on Thermometry for Perioperative Medicine.

Authors:  Andre van Zundert; Tonchanok Intaprasert; Floris Wiepking; Victoria Eley
Journal:  Healthcare (Basel)       Date:  2022-01-24

Review 3.  Temperature measurements with a temporal scanner: systematic review and meta-analysis.

Authors:  Håkan Geijer; Ruzan Udumyan; Georg Lohse; Ylva Nilsagård
Journal:  BMJ Open       Date:  2016-03-31       Impact factor: 2.692

4.  Evaluation of skin temperature over carotid artery for temperature monitoring in comparison to nasopharyngeal temperature in adults under general anesthesia.

Authors:  Venkatesh Selvaraj; Pughal Vendan Gnanaprakasam
Journal:  Anesth Essays Res       Date:  2016 May-Aug

5.  Comparison of Nasopharyngeal Temperature Measured at Fossa of Rosenmuller and Blindly Inserted Temperature Probe with Esophageal Temperature: A Cross-Sectional Study.

Authors:  Arun Kumar Handigodu Duggappa; Shaji Mathew; Dupati Nikkhil Gupta; Shiyad Muhamed; Pawan Nanjangud; Abhishek Rao Kordcal
Journal:  Anesth Essays Res       Date:  2018 Apr-Jun
  5 in total