Literature DB >> 23219052

Detecting cause of dislocation after total hip arthroplasty by patient-specific four-dimensional motion analysis.

Hidenobu Miki1, Nobuhiko Sugano, Kazuo Yonenobu, Kosuke Tsuda, Maki Hattori, Naoki Suzuki.   

Abstract

BACKGROUND: Dislocation is a major complication after total hip arthroplasty. Prosthesis impingement is considered to be an important cause of dislocation because impingement marks are more frequently found on retrieved cups or liners in patients who have undergone revision surgery because of dislocation (80%-94%) than in those who have undergone reoperation for other reasons (51%-56%). However, it remains a question whether impingement marks are the cause of dislocation or are instead its result. To clarify the issue, it is necessary to confirm noninvasively whether the point of impingement matches the patient's hip position when dislocation occurs.
METHODS: Using four-dimensional patient-specific analysis, we recorded prosthesis impingement in 10 hips with instability after primary total hip arthroplasty when the patients reproduced the dislocation-causing motion.
FINDINGS: We found prosthesis impingement to be related to at least instability in 6 of 10 hips with dislocation after primary total hip arthroplasty and in 4 of 4 hips that underwent revision surgery for recurrent dislocation. All impingements occurred between the anterior wall of the liner and the stem neck in posterior dislocation and between the posterior wall of the liner and the stem neck in anterior dislocation. Revision surgery in 1 of those 4 hips revealed 2 impingement marks on the retrieved liner that closely matched the prosthesis impingement point and the dislocation pathway of the metal head on the liner that were detected earlier during motion analysis.
INTERPRETATION: Prosthesis impingement is an important factor in dislocation after total hip arthroplasty.
Copyright © 2012 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23219052     DOI: 10.1016/j.clinbiomech.2012.11.009

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  5 in total

1.  Does dual-mobility cup geometry affect posterior horizontal dislocation distance?

Authors:  Christopher Heffernan; Samik Banerjee; Jim Nevelos; James Macintyre; Kimona Issa; David C Markel; Michael A Mont
Journal:  Clin Orthop Relat Res       Date:  2014-01-24       Impact factor: 4.176

2.  Soft tissue tension is four times lower in the unstable primary total hip arthroplasty.

Authors:  Takeshi Ogawa; Masaki Takao; Hidetoshi Hamada; Takashi Sakai; Nobuhiko Sugano
Journal:  Int Orthop       Date:  2018-03-27       Impact factor: 3.075

3.  Unexpected dislocation following accurate total hip arthroplasty caused by excessive hip joint laxity during myasthenic crisis: a case report.

Authors:  Yoshiki Murotani; Yutaka Kuroda; Koji Goto; Toshiyuki Kawai; Shuichi Matsuda
Journal:  J Med Case Rep       Date:  2018-11-06

4.  Bone-to-Bone and Implant-to-Bone Impingement: A Novel Graphical Representation for Hip Replacement Planning.

Authors:  Arnab Palit; Richard King; Zoe Hart; Yolanda Gu; James Pierrepont; Mark T Elliott; Mark A Williams
Journal:  Ann Biomed Eng       Date:  2020-01-22       Impact factor: 3.934

5.  Long-term Results of the Thrust Plate Prosthesis, 21-year Follow-up: A Case Report and Literature Review.

Authors:  Hany Elbardesy; Rehan Gul
Journal:  J Orthop Case Rep       Date:  2021-04
  5 in total