Literature DB >> 23216124

Development and validation of an ultra-sensitive method for the measurement of plasma renin activity in human plasma via LC-MS/MS.

Derek L Chappell1, Thomas McAvoy, Barbra Weiss, Russell Weiner, Omar F Laterza.   

Abstract

BACKGROUND: Renin catalyzes the conversion of angiotensinogen to angiotensin I (Ang I), the first and rate-limiting step in the renin-angiotensin-aldosterone system. Plasma renin activity (PRA) is an important target engagement biomarker in the clinical development of renin inhibitors. We have developed and validated an improved PRA assay that incorporates an Ang I trapping antibody followed by extraction and quantification of Ang I using a highly sensitive and specific LC-MS/MS method.
RESULTS: The following assay performance characteristics were assessed as part of analytical validation: precision, LOQ, spike recovery, dilution linearity, stability, absolute recovery and biological variability. The assay demonstrated excellent performance characteristics. Notably, the sensitivity of the assay was increased 140-fold when compared with a previous enzyme immunoassay-based assay.
CONCLUSION: The improved sensitivity allowed the measurement of >95% PRA inhibition from baseline levels. In addition, we compared the LC-MS/MS-based assay to an enzyme immunoassay-based assay.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23216124     DOI: 10.4155/bio.12.268

Source DB:  PubMed          Journal:  Bioanalysis        ISSN: 1757-6180            Impact factor:   2.681


  3 in total

1.  Development of a novel nanoflow liquid chromatography-parallel reaction monitoring mass spectrometry-based method for quantification of angiotensin peptides in HUVEC cultures.

Authors:  Chuan He; Simiao Hu; Wanxing Zhou
Journal:  PeerJ       Date:  2020-09-15       Impact factor: 2.984

2.  UPLC-MRM Mass Spectrometry Method for Measurement of the Coagulation Inhibitors Dabigatran and Rivaroxaban in Human Plasma and Its Comparison with Functional Assays.

Authors:  Joachim Kuhn; Tatjana Gripp; Tobias Flieder; Marcus Dittrich; Doris Hendig; Jessica Busse; Cornelius Knabbe; Ingvild Birschmann
Journal:  PLoS One       Date:  2015-12-23       Impact factor: 3.240

3.  Customisation and validation of a low-volume plasma renin activity immunoassay: Enabling of regulatory compliant determination in paediatric trials.

Authors:  F K Suessenbach; J Tins; B B Burckhardt
Journal:  Pract Lab Med       Date:  2019-11-12
  3 in total