Determination of creatinine in human urine with flow injection tandem mass spectrometry.

Mareile Niesser1, Berthold Koletzko, Wolfgang Peissner.   

Abstract

BACKGROUND/AIMS: Excretion of urinary compounds in spot urine is often estimated relative to creatinine. For the growing number of liquid chromatography-tandem mass spectrometry (LC-MS/MS) assays of urine-excreted molecules, a fast and accurate method for determination of creatinine is needed.
METHODS: A high-throughput flow injection tandem mass spectrometry method for exact quantitation of creatinine in urine has been developed and validated. Sample preparation used only two-step dilution for protein precipitation and matrix dilution. Flow injection analysis without chromatographic separation allowed for total run times of 1 min per sample. Creatinine concentrations were quantitated using stable isotope dilution tandem mass spectrometry. Selectivity and coelution-free quantitation were assured by qualifier ion monitoring.
RESULTS: Method validation revealed excellent injection repeatability of 1.0% coefficient of variation (CV), intraday precision of 1.2% CV and interday precision of 2.4% CV. Accuracy determined from standard addition experiments was 106.1 ± 3.8%. The linear calibration range was adapted to physiological creatinine concentrations. Comparison of quantitation results with a routinely used method (Jaffé colorimetric assay) proved high agreement (R(2) = 0.9102).
CONCLUSIONS: The new method is a valuable addition to the toolbox of LC-MS/MS laboratories where excretion of urinary compounds is studied. The 'dilute and shoot' approach to isotope dilution tandem mass spectrometry makes the new method highly accurate as well as cost- and time-efficient.
Copyright © 2012 S. Karger AG, Basel.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23208161     DOI: 10.1159/000342774

Source DB:  PubMed          Journal:  Ann Nutr Metab        ISSN: 0250-6807            Impact factor:   3.374


  4 in total

1.  Albumin testing in urine using a smart-phone.

Authors:  Ahmet F Coskun; Richie Nagi; Kayvon Sadeghi; Stephen Phillips; Aydogan Ozcan
Journal:  Lab Chip       Date:  2013-11-07       Impact factor: 6.799

2.  Urinary metabolites in patients undergoing coronary catheterization via the radial versus femoral artery approach.

Authors:  Anupama Vasudevan; Jeffrey M Schussler; Jane I Won; Paula Ashcraft; Ivy Bolanos; Matthew Williams; Teodoro Bottiglieri; Carlos E Velasco; Peter A McCullough
Journal:  Proc (Bayl Univ Med Cent)       Date:  2017-10

3.  A highly sensitive liquid chromatography electrospray ionization mass spectrometry method for quantification of TMA, TMAO and creatinine in mouse urine.

Authors:  Sunil Veeravalli; Kersti Karu; Ian R Phillips; Elizabeth A Shephard
Journal:  MethodsX       Date:  2017-09-28

4.  Sensitivity Detection of Uric Acid and Creatinine in Human Urine Based on Nanoporous Gold.

Authors:  Keshuai Shang; Shuangjue Wang; Siyu Chen; Xia Wang
Journal:  Biosensors (Basel)       Date:  2022-08-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.