| Literature DB >> 23207656 |
Ludomir Stefańczyk1, Marcin T Elgalal, Wojciech Szubert, Piotr Grzelak, Piotr Szopiński, Agata Majos.
Abstract
A case of femoral artery obstruction following application of a StarClose type arterial puncture closing device (APCD) is presented. Ultrasonographic and angiographic imaging of this complication was obtained. The posterior wall of the vessel was accidentally caught in the anchoring element of the nitinol clip. This complication was successfully resolved by endovascular treatment and the implantation of a stent.Entities:
Mesh:
Substances:
Year: 2012 PMID: 23207656 DOI: 10.1007/s00270-012-0522-x
Source DB: PubMed Journal: Cardiovasc Intervent Radiol ISSN: 0174-1551 Impact factor: 2.740