Increased intraosseous temperature caused by ultrasonic devices during bone surgery and the influences of working pressure and cooling irrigation.

Falk Birkenfeld1, Merlind Erika Becker, Sönke Harder, Ralph Lucius, Matthias Kern.   

Abstract

PURPOSE: The purpose of this study was to investigate the increases in intraosseous temperature generated by a modern ultrasonic device for bone surgery (UDBS) and the influences of working pressure and cooling irrigation on this temperature.
MATERIALS AND METHODS: Twenty human mandibular bone specimens (20x15x5 to 7 mm) were used; three vertical cuts were performed for a duration of 12 seconds per cut. Each bone specimen was machined with a different combination of working pressure (1.5, 2.0, 3.0, 4.0, or 6.0 N) and cooling irrigation (0, 30, 60, or 90 mL/min), and intraosseous temperatures were measured. Harmful temperature development was defined as an increase of more than 10°C for the 75th percentile and/or a maximum increase of more than 15°C. Cutting performance was also measured.
RESULTS: Harmless intraosseous temperature development was identified for working pressures of 1.5 N and 2.0 N with cooling irrigations of 30, 60, and 90 mL/min and for 3.0 N at 90 mL/min. The maximum temperature observed was 72°C (6.0 N with 60 mL/min). The mean cutting performance values were 0.21±0.02 mm/s for 6.0 N, 0.21±0.06 mm/s for 3.0 N, 0.20±0.01 mm/s for 4.0 N, 0.11±0.05 mm/s for 1.5 N, and 0.08±0.03 mm/s for 2.0 N.
CONCLUSIONS: To prevent tissue damage in dental bone surgery, a minimum coolant amount of 30 mL/min is recommended. The working pressure should be chosen with great care because of its significant influence on intraosseous temperature. Doubling of the working pressure from 1.5 to 3.0 N requires a tripling of the coolant (30 to 90 mL/min) to prevent tissue damage. A working pressure above 3.0 N did not result in improved cutting performance.

Entities:  

Mesh:

Year:  2012        PMID: 23189287

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  3 in total

1.  Influence of surgical ultrasound used in the detachment of flaps, osteotomy and odontosection in lower third molar surgeries. A prospective, randomized, and "split-mouth" clinical study.

Authors:  L-D Silva; E-N Reis; J-P Bonardi; V-N Lima; A-M Aranega; D Ponzoni
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2020-07-01

Review 2.  Piezosurgery in Hallux Valgus Correction: Distal Linear Osteotomy Operative Technique Using Piezoelectric Tools.

Authors:  Alessandro Russo; Silvio Caravelli; Massimiliano Mosca; Mauro Girolami; Alessandro Ortolani; Simone Massimi; Mario Fuiano; Stefano Zaffagnini
Journal:  Joints       Date:  2019-12-13

Review 3.  Piezoelectric Bone Surgery: A Review of the Literature and Potential Applications in Veterinary Oromaxillofacial Surgery.

Authors:  Philippe Hennet
Journal:  Front Vet Sci       Date:  2015-05-05
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.