Indium-111 DTPA flow study to evaluate surgically implanted drug pump delivery system.

A S Rosenson1, A Ali, E W Fordham, R D Penn.   

Abstract

The authors describe a radionuclide technique for evaluating intrathecal and intraventricular infusions by surgically implanted drug pump delivery systems. Sixteen patients underwent flow studies, performed by injecting 500 mu Ci of indium-111 DTPA into the pumps. Early and delayed images were obtained. These studies enabled distinction among functioning pumps, nonfunctioning pumps, and obstructed or occluded catheters. We conclude that indium-111 DTPA flow studies provide an excellent way to assess these drug pump delivery systems.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2317993     DOI: 10.1097/00003072-199003000-00003

Source DB:  PubMed          Journal:  Clin Nucl Med        ISSN: 0363-9762            Impact factor:   7.794


  2 in total

Review 1.  Imaging evaluation of intrathecal baclofen pump-catheter systems.

Authors:  A C Miracle; M A Fox; R N Ayyangar; A Vyas; S K Mukherji; D J Quint
Journal:  AJNR Am J Neuroradiol       Date:  2010-10-28       Impact factor: 3.825

2.  Evaluation of usefulness of scintigraphic imaging in diagnosis of intrathecal drug delivery system malfunction - a preliminary report.

Authors:  Jacek Teodorczyk; Tomasz Szmuda; Mariusz Siemiński; Piotr Lass; Paweł Słoniewski
Journal:  Pol J Radiol       Date:  2013-07
  2 in total