Literature DB >> 2317933

A microalbuminuria assay using bromphenol blue.

K Jung1, E Nickel, M Pergande.   

Abstract

A simple microalbuminuria assay using bromphenol blue/glycine reagent is described. Urine samples were prepared using gel filtration on Sephadex G-50 minicolumns and absorbance was measured at 610 nm 20 s after mixing 10 parts of eluate and 1 part of reagent. The detection limit of this method was 3 mg/l; within-run and between-run precision was between 0.5 and 4.1% for borderline and raised albumin concentrations. The recovery of albumin added to samples was 98.7 +/- 2.5%. Results obtained by this method correlated closely with values obtained by radial immunodiffusion (r = 0.987). The test is cheap (reagent costs about 5 cents) and suitable for the non-specialist laboratory.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2317933     DOI: 10.1016/0009-8981(90)90343-q

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  1 in total

1.  Porous polymer coatings as substrates for the formation of high-fidelity micropatterns by quill-like pens.

Authors:  Michael Hirtz; Marcus Lyon; Wenqian Feng; Andrea E Holmes; Harald Fuchs; Pavel A Levkin
Journal:  Beilstein J Nanotechnol       Date:  2013-06-19       Impact factor: 3.649

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.