Literature DB >> 23174063

Biomechanical comparison of parallel versus 90-90 plating of bicolumn distal humerus fractures with intra-articular comminution.

Christopher Got1, John Shuck, Alison Biercevicz, Dave Paller, Mary Mulcahey, Matthew Zimmermann, Theodore Blaine, Andrew Green.   

Abstract

PURPOSE: To compare the biomechanical properties of 90-90 versus mediolateral parallel plating of C-3 bicolumn distal humerus fractures.
METHODS: We created intra-articular AO/Orthopaedic Trauma Association C-3 bicolumn fractures in 10 fresh-frozen matched pairs of cadaveric elbows. We determined bone mineral density of the metaphyseal region with dual-energy x-ray absorptiometry. The matched pairs of elbows were randomly assigned to either 90-90 or parallel plate fixation. We tested anteroposterior displacement at a rate of 0.5 mm/s to a maximum load of ± 100 N for both the articular and entire distal humerus segments. We tested torsional stability at a displacement rate of 0.1 Hz to a maximum torque of ± 2.5 Nm. After cyclical testing, we loaded the specimens in torsion to failure.
RESULTS: There was no significant difference in the bone density of the paired specimens. Compared with parallel fixation, 90-90 plate fixation had significantly greater torque to failure load. Both plating constructs were equally sensitive to bone density. Both techniques had the same mode of failure in torsion, a spiral fracture extending from the medial plate at the metaphyseal-diaphyseal junction. There was no significant difference in the stiffness of fixation of the articular fragment or the entire distal segment in anteroposterior loading.
CONCLUSIONS: This study demonstrated that 90-90 and parallel plating had comparable biomechanical properties for fixation of comminuted intra-articular distal humerus fractures, and that 90-90 plating had greater resistance to torsional loading.
Copyright © 2012 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23174063     DOI: 10.1016/j.jhsa.2012.08.042

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  7 in total

1.  Bicolumnar 90-90 plating of low-energy distal humeral fractures in the elderly patient.

Authors:  Daniel F Leigey; Dana J Farrell; Peter A Siska; Ivan S Tarkin
Journal:  Geriatr Orthop Surg Rehabil       Date:  2014-09

Review 2.  Intra-articular fractures of the distal humerus-a review of the current practice.

Authors:  Charalampos G Zalavras; Efthymios Papasoulis
Journal:  Int Orthop       Date:  2018-02-05       Impact factor: 3.075

3.  Parallel versus orthogonal plate osteosynthesis of adult distal humerus fractures: a meta-analysis of biomechanical studies.

Authors:  Chien-An Shih; Wei-Ren Su; Wei-Chin Lin; Tai-Wei Tai
Journal:  Int Orthop       Date:  2018-04-20       Impact factor: 3.075

4.  A comparison between orthogonal and parallel plating methods for distal humerus fractures: a prospective randomized trial.

Authors:  Sang Ki Lee; Kap Jung Kim; Kyung Hoon Park; Won Sik Choy
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-08-07

5.  Bicolumnar 90-90 plating of AO 13C type fractures.

Authors:  Cemal Kural; Ersin Ercin; Mehmet Erkilinc; Evren Karaali; Mustafa Gokhan Bilgili; Suleyman Altun
Journal:  Acta Orthop Traumatol Turc       Date:  2017-03-22       Impact factor: 1.511

6.  Biomechanical Evaluation of Dual Plate Configurations for Femoral Shaft Fracture Fixation.

Authors:  Marc El Beaino; Randal P Morris; Ronald W Lindsey; Zbigniew Gugala
Journal:  Biomed Res Int       Date:  2019-04-28       Impact factor: 3.411

7.  Results of plate fixation for transcondylar fracture of the distal humerus: a rare pattern of fractures.

Authors:  Yoo Joon Sur; Yoochang Kim; Ho Youn Park
Journal:  JSES Int       Date:  2020-05-29
  7 in total