The accuracy of ultrasound volume flow measurements in the complex flow setting of a forearm vascular access.

Koen Van Canneyt1, Abigail Swillens, Lasse Lovstakken, Luca Antiga, Pascal Verdonck, Patrick Segers.   

Abstract

PURPOSE: Maturation of an arterio-venous fistula (AVF) frequently fails, with low post-operative fistula flow as a prognostic marker for this event. As pulsed wave Doppler (PWD) is commonly used to assess volume flow, we studied the accuracy of this measurement in the setting of a radio-cephalic AVF.
METHODS: As in-vivo validation of fistula flow measurements is cumbersome, we performed simulations, integrating computational fluid dynamics with an ultrasound (US) simulator. Flow in the arm was calculated, based on a patient-specific model of the arm vasculature pre and post AVF creation. Raw ultrasound signals were subsequently simulated, from which Doppler spectra were calculated in both a proximal and a distal location.
RESULTS: The velocity component in the direction of the PWD-US beam (vPWD), in a centered, small, sample volume, can be captured accurately using PWD spectrum mean-tracking (maximum bias [mB] 8.1%). However, when deriving flow rate from these measurements, a high degree of inaccuracy occurs. First, the angle-correction of vPWD towards the velocity along the axis of the vessel is largely influenced by the radial velocity components in the complex flow field (mB=16.3%). Second, the largest error is introduced when transferring the centerline velocity to the cross-sectional mean velocity without any knowledge of the flow profile (mB=97.7%).
CONCLUSIONS: In the setting of a forearm AVF, flow estimates based on PWD are hampered by the complex flow patterns. Overall, flow estimation based on centerline measurement, analyzed by mean-tracking of the RF-spectral estimates, under the assumption of a parabolic flow profile, appeared to provide the most reasonable values.

Entities:  

Mesh:

Year:  2012        PMID: 23172170     DOI: 10.5301/jva.5000118

Source DB:  PubMed          Journal:  J Vasc Access        ISSN: 1129-7298            Impact factor:   2.283


  3 in total

1.  Jugular Venous Flow Quantification Using Doppler Sonography.

Authors:  Karen Marr; Dejan Jakimovski; Marcello Mancini; Ellen Carl; Robert Zivadinov
Journal:  Ultrasound Med Biol       Date:  2018-05-18       Impact factor: 2.998

2.  Assessment of Spectral Doppler for an Array-Based Preclinical Ultrasound Scanner Using a Rotating Phantom.

Authors:  David A Kenwright; Tom Anderson; Carmel M Moran; Peter R Hoskins
Journal:  Ultrasound Med Biol       Date:  2015-05-06       Impact factor: 2.998

3.  An in vitro Assessment of the Haemodynamic Features Occurring Within the True and False Lumens Separated by a Dissection Flap for a Patient-Specific Type B Aortic Dissection.

Authors:  Liam Morris; Paul Tierney; Niamh Hynes; Sherif Sultan
Journal:  Front Cardiovasc Med       Date:  2022-03-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.