The effect of implant design of linked total elbow arthroplasty on stability and stress: a finite element analysis.

Ryan Willing1, Graham J W King, James A Johnson.   

Abstract

Several linked total elbow arthroplasty designs exist, which function similar to a loose hinge joint. Constraint behaviour is an important design consideration, as it affects joint stability, or how much secondary [e.g. varus-valgus (VV)] motion is permitted. Implant durability is also a concern, as bearing failures have been reported. This finite element analysis investigates the constraint characteristics and ultra high molecular weight polyethylene bearing stresses of three linked elbow design concepts [cylindrical (CY), hourglass (HG) and concave cylinder (CC)]. The bearing of the CY design was subjected to elevated Von Mises stresses (2.1-5.4 times higher than the HG and CC designs) due to edge loading. The HG design maintained low stresses, but was unable to provide consistent VV stability. The CC design also maintained low stresses while providing consistent VV stability. These results suggest that CC designs may provide better stability characteristics and durability in vivo, compared to the other two designs.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23171038     DOI: 10.1080/10255842.2012.739161

Source DB:  PubMed          Journal:  Comput Methods Biomech Biomed Engin        ISSN: 1025-5842            Impact factor:   1.763


  3 in total

Review 1.  What design and material factors impact the wear and corrosion performance in total elbow arthroplasties?

Authors:  Mark P Figgie; Timothy M Wright; Denise Drinkwater
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

Review 2.  [Elbow prosthesis after acute fractures : Indications and technique].

Authors:  T Leschinger; M Hackl; F Lanzerath; F Krane; A Harbrecht; K Wegmann; L P Müller
Journal:  Unfallchirurgie (Heidelb)       Date:  2022-07-14

3.  Primary total elbow arthroplasty.

Authors:  Armin Badre; Graham J W King
Journal:  J Clin Orthop Trauma       Date:  2021-04-20
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.