An easy solution to obstructed sampling line: Auxillary oxygen flowmeter.

Swati Chhabra1, Susheela Taxak.   

Abstract

Entities:  

Year:  2012        PMID: 23162420      PMCID: PMC3498685          DOI: 10.4103/1658-354X.101239

Source DB:  PubMed          Journal:  Saudi J Anaesth


× No keyword cloud information.
Sir, This is with reference to the correspondence by Nataraj Madagondapalli Srinivasan, “Use of heat and moisture exchanger in intubated patients reduces blockage in gas sampling tube of the mass spectrometer,”[1] in the Jan–Mar 2011 issue of the Saudi Journal of Anaesthesia. We read it with interest as it discusses a rather common problem. Capnography is being widely used intraoperatively for confirmation of endotracheal tube placement and for adequacy of ventilation during anesthesia. If facilities are available, the gas sampled for capnography is also used for anesthetic gas monitoring using the principle of mass spectrometry. Obstruction of sampling line due to water droplets or secretions[2] is not infrequent, especially during long procedures or when a single sampling line is used for a number of cases. Although the water trap prevents entry of these droplets into the monitor, tubing obstruction can lead to erroneous readings or no readings at all. Many methods are described in the literature to get rid of the obstruction, viz. high gas flows, flushing with a syringe and heat and moisture exchanger.[1] Recently, during a procedure over the head and neck, we encountered an obstruction such that our anesthesia gas monitor could not analyze the sampled gas to give us readings. The distal end of the sampling line was under the drapes and thus not accessible. The proximal end was not compatible for syringe attachment [Figure 1]. Therefore, we connected the proximal end to the auxillary oxygen flowmeter of the anesthesia machine with oxygen flow at 7 L/min to flush out the obstruction [Figure 2].
Figure 1

Incompatibility of syringe and sampling line

Figure 2

Proximal end of sampling line connected to auxillary oxygen flowmeter

Incompatibility of syringe and sampling line Proximal end of sampling line connected to auxillary oxygen flowmeter When the sampling line was reattached to the monitor [Figure 3], proper capnograph and readings were obtained. Thus, this becomes an effective method of relieving the sampling line obstruction, which does not require any additional equipment. Although there is a disadvantage of dilution of anesthetic gases, it is very transient as high oxygen flows are used for a short duration of time.
Figure 3

AGM with water trap and proximal sampling line

AGM with water trap and proximal sampling line
  2 in total

1.  Erroneous mass spectrometer data caused by a faulty patient sampling tube: case report and laboratory study.

Authors:  T M Skeehan; J F Biebuyck
Journal:  J Clin Monit       Date:  1991-10

2.  Use of heat and moisture exchanger in intubated patients reduces the blockage in gas sampling tube of the mass spectrometer.

Authors:  Nataraj Madagondapalli Srinivasan
Journal:  Saudi J Anaesth       Date:  2011-01
  2 in total
  3 in total

1.  An indigenous method to prevent intraoperative kinking of gas sampling line.

Authors:  Lenin Babu Elakkumanan; Sivaraman Baskaran; Senthilnathan Muthapillai
Journal:  Indian J Anaesth       Date:  2013-11

2.  Yet another way to clear water drops from ETCO2 sampling line.

Authors:  Akshaya N Shetti
Journal:  Saudi J Anaesth       Date:  2014-07

3.  A method to prevent kinking of gas sampling line.

Authors:  Akshaya N Shetti; Sangamesh Kunakeri; Shivanand L Karigar
Journal:  Anesth Essays Res       Date:  2014 May-Aug
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.