Literature DB >> 23160241

Sodium citrate vacuum tubes validation: preventing preanalytical variability in routine coagulation testing.

Gabriel Lima-Oliveira1, Giuseppe Lippi, Gian Luca Salvagno, Martina Montagnana, Geraldo Picheth, Gian Cesare Guidi.   

Abstract

Sometimes in-vitro diagnostic devices (e.g. blood collection tubes) are not validated before use or when the producer's brand is changed. The aim of this study was to validate five brands of sodium citrate vacuum tubes. Blood specimens from 50 volunteers were collected in five different tube brands (I: Venosafe, II: VACUETTE, III: BD Vacutainer, IV: LABOR IMPORT and V: S-Monovette). Routine coagulation tests [activated partial thromboplastin time (aPTT), prothrombin time (PT), and fibrinogen (FIB)] were performed on ACL TOP instrument using HemosIL reagents. The significance of the differences between samples was assessed by paired Student's t-test, set at P < 0.005. Significant differences were observed for: PT when comparing I vs. II, I vs. III, I vs. V, II vs. III, II vs. IV, II vs. V, III vs. IV, III vs. V and IV vs. V; aPTT when comparing I vs. II, I vs. III, I vs. IV, II vs. IV, III vs. IV and IV vs. V. No differences were observed among brands for FIB determination. We suggest that every laboratory management should both standardize the procedures and frequently evaluate the quality of in-vitro diagnostic devices.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23160241     DOI: 10.1097/MBC.0b013e32835b72ea

Source DB:  PubMed          Journal:  Blood Coagul Fibrinolysis        ISSN: 0957-5235            Impact factor:   1.276


  7 in total

1.  A new device to relieve venipuncture pain can affect haematology test results.

Authors:  Gabriel Lima-Oliveira; Giuseppe Lippi; Gian Luca Salvagno; Marise Danielle Raulino Campelo; Katharyne Soares Adala Tajra; Flavio dos Santos Gomes; Carlos David Valentim; Sylvio José Colonna Romano; Geraldo Picheth; Gian Cesare Guidi
Journal:  Blood Transfus       Date:  2013-07-18       Impact factor: 3.443

2.  Quality impact on diagnostic blood specimen collection using a new device to relieve venipuncture pain.

Authors:  Gabriel Lima-Oliveira; Giuseppe Lippi; Gian Luca Salvagno; Martina Montagnana; Geraldo Picheth; Gian Cesare Guidi
Journal:  Indian J Clin Biochem       Date:  2013-04-03

3.  Guidance on the critical shortage of sodium citrate coagulation tubes for hemostasis testing.

Authors:  Robert C Gosselin; Annette Bowyer; Emmanuel J Favaloro; Jill M Johnsen; Giuseppe Lippi; Richard A Marlar; Keith Neeves; Marian A Rollins-Raval
Journal:  J Thromb Haemost       Date:  2021-09-02       Impact factor: 16.036

4.  Could light meal jeopardize laboratory coagulation tests?

Authors:  Gabriel Lima-Oliveira; Gian Luca Salvagno; Giuseppe Lippi; Elisa Danese; Matteo Gelati; Martina Montagnana; Geraldo Picheth; Gian Cesare Guidi
Journal:  Biochem Med (Zagreb)       Date:  2014-10-15       Impact factor: 2.313

5.  Stability of 21 Antihypertensive Drugs in Serum Collected in Standard (Nongel) Serum Tubes Versus Tubes Containing a Gel Separator.

Authors:  Solfrid Hegstad; Olav Spigset; Arne Helland
Journal:  Ther Drug Monit       Date:  2020-04       Impact factor: 3.118

6.  The effective reduction of tourniquet application time after minor modification of the CLSI H03-A6 blood collection procedure.

Authors:  Gabriel Lima-Oliveira; Giuseppe Lippi; Gian Luca Salvagno; Martina Montagnana; Geraldo Picheth; Gian Cesare Guidi
Journal:  Biochem Med (Zagreb)       Date:  2013       Impact factor: 2.313

7.  Incorrect order of draw could be mitigate the patient safety: a phlebotomy management case report.

Authors:  Gabriel Lima-Oliveira; Giuseppe Lippi; Gian Luca Salvagno; Martina Montagnana; Geraldo Picheth; Gian Cesare Guidi
Journal:  Biochem Med (Zagreb)       Date:  2013       Impact factor: 2.313

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.