Pacemaker malfunction due to mechanical failure of the lead-header interface.

Amanda E Coleman1, Teresa C Defrancesco, Guillaume Chanoit.   

Abstract

An 8 year old female spayed Boxer dog, diagnosed with concurrent vasovagal syncope and arrhythmogenic right ventricular cardiomyopathy, presented for routine evaluation approximately three months following epicardial pacemaker implantation. Routine device interrogation revealed intermittent loss of ventricular capture and intermittent failure to appropriately sense. Following evaluation of chronic impedance data, failure of the pacemaker lead-header interface or lead conductor fracture was suspected. Radiographic and pacemaker interrogator findings suggested incomplete lead insertion into the device header with intermittent loss of ventricular capture and variable pacemaker sensing. We hypothesize that either the presence of a loose or cross-threaded set screw or operator error at the time of device implantation may have caused this complication. This report details the diagnosis of mechanical failure of the lead-header interface, a complication not previously reported in a veterinary patient.
Copyright © 2012 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23158383     DOI: 10.1016/j.jvc.2012.07.003

Source DB:  PubMed          Journal:  J Vet Cardiol        ISSN: 1760-2734            Impact factor:   1.701


  1 in total

1.  A retrospective review of 146 active and passive fixation bradycardia lead implantations in 74 dogs undergoing pacemaker implantation in a research setting of short term duration.

Authors:  Lynne E Swanson; Barbara A Huibregtse; Brian A Scansen
Journal:  BMC Vet Res       Date:  2018-03-27       Impact factor: 2.741

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.