Influence of total beam current on HRTEM image resolution in differentially pumped ETEM with nitrogen gas.

A N Bright1, K Yoshida, N Tanaka.   

Abstract

Environmental transmission electron microscopy (ETEM) enables the study of catalytic and other reaction processes as they occur with Angstrom-level resolution. The microscope used is a dedicated ETEM (Titan ETEM, FEI Company) with a differential pumping vacuum system and apertures, allowing aberration corrected high-resolution transmission electron microscopy (HRTEM) imaging to be performed with gas pressures up to 20 mbar in the sample area and with significant advantages over membrane-type E-cell holders. The effect on image resolution of varying the nitrogen gas pressure, electron beam current density and total beam current were measured using information limit (Young's fringes) on a standard cross grating sample and from silicon crystal lattice imaging. As expected, increasing gas pressure causes a decrease in HRTEM image resolution. However, the total electron beam current also causes big changes in the image resolution (lower beam current giving better resolution), whereas varying the beam current density has almost no effect on resolution, a result that has not been reported previously. This behavior is seen even with zero-loss filtered imaging, which we believe shows that the drop in resolution is caused by elastic scattering at gas ions created by the incident electron beam. Suitable conditions for acquiring high resolution images in a gas environment are discussed. Lattice images at nitrogen pressures up to 16 mbar are shown, with 0.12 nm information transfer at 4 mbar.
Copyright © 2012 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23142744     DOI: 10.1016/j.ultramic.2012.08.007

Source DB:  PubMed          Journal:  Ultramicroscopy        ISSN: 0304-3991            Impact factor:   2.689


  2 in total

1.  Assessing and ameliorating the influence of the electron beam on carbon nanotube oxidation in environmental transmission electron microscopy.

Authors:  Ai Leen Koh; Robert Sinclair
Journal:  Ultramicroscopy       Date:  2016-12-10       Impact factor: 2.689

2.  Water without windows: Evaluating the performance of open cell transmission electron microscopy under saturated water vapor conditions, and assessing its potential for microscopy of hydrated biological specimens.

Authors:  Cathal Cassidy; Masao Yamashita; Martin Cheung; Chola Kalale; Hidehito Adaniya; Ryusuke Kuwahara; Tsumoru Shintake
Journal:  PLoS One       Date:  2017-11-03       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.