ProDisc cervical arthroplasty does not alter facet joint contact pressure during lateral bending or axial torsion.

Nicolas V Jaumard1, Joel A Bauman, Benjamin B Guarino, Akhilesh J Gokhale, Daniel E Lipschutz, Christine L Weisshaar, William C Welch, Beth A Winkelstein.   

Abstract

STUDY
DESIGN: A biomechanical study of facet joint pressure after total disc replacement using cadaveric human cervical spines during lateral bending and axial torsion.
OBJECTIVE: The goal was to measure the contact pressure in the facet joint in cadaveric human cervical spines subjected to physiologic lateral bending and axial torsion before and after implantation of a ProDisc-C implant. SUMMARY OF BACKGROUND DATA: Changes in facet biomechanics can damage the articular cartilage in the joint, potentially leading to degeneration and painful arthritis. Few cadaveric and computational studies have evaluated the changes in facet joint loading during spinal loading with an artificial disc implanted. Computational models have predicted that the design and placement of the implant influence facet joint loading, but limited cadaveric studies document changes in facet forces and pressures during nonsagittal bending after implantation of a ProDisc. As such, little is known about the local facet joint mechanics for these complicated loading scenarios in the cervical spine.
METHODS: Seven osteoligamentous C2-T1 cadaveric cervical spines were instrumented with a transducer to measure the C5-C6 facet pressure profiles during physiological lateral bending and axial torsion, before and after implantation of a ProDisc-C at that level. Rotations at that level and global cervical spine motions and loads were also quantified. RESULT.: Global and segmental rotations were not altered by the disc implantation. Facet contact pressure increased after implantation during ipsilateral lateral bending and contralateral torsion, but that increase was not significant compared with the intact condition.
CONCLUSION: Implantation of a ProDisc-C does not significantly modify the kinematics and facet pressure at the index level in cadaveric specimens during lateral bending and axial torsion. However, changes in facet contact pressures after disc arthroplasty may have long-term effects on spinal loading and cartilage degeneration and should be monitored in vivo.

Entities:  

Mesh:

Year:  2013        PMID: 23132537     DOI: 10.1097/BRS.0b013e31827b8a2d

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  4 in total

Review 1.  Cervical disc replacement surgery: biomechanical properties, postoperative motion, and postoperative activity levels.

Authors:  Alfred Pisano; Melvin Helgeson
Journal:  Curr Rev Musculoskelet Med       Date:  2017-06

2.  Load rate of facet joints at the adjacent segment increased after fusion.

Authors:  Hui Li; Bao-Qing Pei; Jin-Cai Yang; Yong Hai; De-Yu Li; Shu-Qin Wu
Journal:  Chin Med J (Engl)       Date:  2015-04-20       Impact factor: 2.628

3.  Anterior bone loss after cervical disc replacement: A systematic review.

Authors:  Xiao-Fei Wang; Yang Meng; Hao Liu; Ying Hong; Bei-Yu Wang
Journal:  World J Clin Cases       Date:  2020-11-06       Impact factor: 1.337

4.  Facet joint disturbance induced by miniscrews in plated cervical laminoplasty: Dose it influence the clinical and radiologic outcomes?

Authors:  Hua Chen; Huibo Li; Beiyu Wang; Tao Li; Quan Gong; Yueming Song; Hao Liu
Journal:  Medicine (Baltimore)       Date:  2016-09       Impact factor: 1.889

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.