| Literature DB >> 23126746 |
F Bencivenga1, A Battistoni, D Fioretto, A Gessini, J R Sandercock, C Masciovecchio.
Abstract
We report on a high resolution inelastic UV scattering table-top setup conceived for Brillouin measurements. The system is based on a tandem 1+1 pass scanning Fabry-Perot interferometer of Sandercock type. Special optics were used in order to adapt such an interferometric device, nowadays only used at visible or IR wavelength, to the UV range. The advantages with respect to other UV Brillouin scattering instruments are the larger resolving power and the improved contrast in the low frequency spectral region. To corroborate these features we provide a comparison between data obtained using the described system and those from existing UV Brillouin scattering instruments.Year: 2012 PMID: 23126746 DOI: 10.1063/1.4756690
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523