Literature DB >> 23124492

Treatment of periprosthetic tibial plateau fractures in unicompartmental knee arthroplasty: plates versus cannulated screws.

Joern Bengt Seeger1, S Jaeger, E Röhner, H Dierkes, G Wassilew, M Clarius.   

Abstract

INTRODUCTION: Periprosthetic tibial plateau fractures (TPF) are rare but represent a serious complication in unicompartmental knee arthroplasty. The most common treatment for these fractures is osteosynthesis with cannulated screws or plates. The aim of this study was to evaluate two different treatment options for periprosthetic fractures. The hypothesis was that angle-stable plates show significantly higher fracture loads than fixation with cannulated screws.
MATERIALS AND METHODS: Twelve matched, paired fresh-frozen tibiae with periprosthetic TPF were used for this study. In Group A, osteosyntheses with cannulated screws were performed, whereas in Group B plates fixated the periprosthetic fracture. DEXA bone density measurement and standard X-rays (AP and lateral) were performed before loading the tibiae under standardised conditions with a maximum load of up to 10.0 kN. After the specimens had been loaded, fracture patterns and fracture loads were analysed and correlated with BMD, BMI, bodyweight (BW), age and size of the tibial implant.
RESULTS: In the plate group all tibiae fracture occured with a median load of F (max) = 2.64 (0.45-5.68) kN, whereas in the group with cannulated screws fractures occurred at a mean load of F (max) = 1.50 (0.27-3.51) kN. The difference was statistically significant at p < 0.05. DISCUSSION: Angle-stable plates showed significantly higher fracture loads than fixation with cannulated screws. Cannulated screws show a reduced stability of the tibial plateau. Therefore in periprosthetic TPF, osteosyntheses with angle-stable plates should be recommended instead of cannulated screws.

Entities:  

Mesh:

Year:  2012        PMID: 23124492     DOI: 10.1007/s00402-012-1649-6

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  4 in total

Review 1.  Periprosthetic fractures after medial unicompartmental knee arthroplasty: a narrative review.

Authors:  L Thoreau; D Morcillo Marfil; E Thienpont
Journal:  Arch Orthop Trauma Surg       Date:  2021-07-15       Impact factor: 2.928

Review 2.  Unicompartmental Knee Arthroplasty: Modes of Failure and Conversion to Total Knee Arthroplasty.

Authors:  Michele Vasso; Katia Corona; Rocco D'Apolito; Giuseppe Mazzitelli; Alfredo Schiavone Panni
Journal:  Joints       Date:  2017-06-05

3.  Periprosthetic tibial fractures in total knee arthroplasty - an outcome analysis of a challenging and underreported surgical issue.

Authors:  Anna Janine Schreiner; Florian Schmidutz; Atesch Ateschrang; Christoph Ihle; Ulrich Stöckle; Björn Gunnar Ochs; Christoph Gonser
Journal:  BMC Musculoskelet Disord       Date:  2018-09-11       Impact factor: 2.362

Review 4.  Comparable incidence of periprosthetic tibial fractures in cementless and cemented unicompartmental knee arthroplasty: a systematic review and meta-analysis.

Authors:  Joost A Burger; Tjeerd Jager; Matthew S Dooley; Hendrik A Zuiderbaan; Gino M M J Kerkhoffs; Andrew D Pearle
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2021-02-02       Impact factor: 4.342

  4 in total