Literature DB >> 23123997

Transapical 29-mm Edwards SAPIEN-XT aortic valve in a 34-mm mitral annuloplasty ring.

Gry Dahle1, Arnt E Fiane, Kjell-Arne Rein.   

Abstract

A 71-year-old woman with severe congestive heart failure and failed mitral valve repair was referred for surgery. Because of her low ejection fraction, a valve-in-ring procedure was suggested. There was a great difference in the given size for the 34-mm Carpentier-Edwards-Physio-Ring and the biggest available transcatheter valve of 29 mm from Edwards. Therefore, we did a "bench test." We expanded a 29-mm Edwards SAPIEN-XT aortic valve in a 34-mm Carpentier-Edwards-Physio-Ring. It fitted well and turned out circular with good coaptation of the leaflets. Thereafter, a successful transapical mitral valve-in-ring implantation on cardiopulmonary bypass was performed, and additional leads for cardiac resynchronization was placed.

Entities:  

Mesh:

Year:  2012        PMID: 23123997     DOI: 10.1097/IMI.0b013e31826d5cd0

Source DB:  PubMed          Journal:  Innovations (Phila)        ISSN: 1556-9845


  2 in total

1.  Update of transcatheter valve treatment.

Authors:  Xian-bao Liu; Jian-an Wang
Journal:  J Zhejiang Univ Sci B       Date:  2013-08       Impact factor: 3.066

Review 2.  Transcatheter mitral valve implantation for degenerated mitral bioprostheses or failed surgical annuloplasty rings: A systematic review and meta-analysis.

Authors:  Junjie Hu; Yan Chen; Sijin Cheng; San Zhang; Kaiqin Wu; Wenli Wang; Yongxin Zhou
Journal:  J Card Surg       Date:  2018-07-10       Impact factor: 1.620

  2 in total