Effects of intra-aortic balloon pump timing on baroreflex activities in a closed-loop cardiovascular hybrid model.

Libera Fresiello1, Ashraf William Khir, Arianna Di Molfetta, Maciej Kozarski, Gianfranco Ferrari.   

Abstract

Despite 50 years of research to assess the intra-aortic balloon pump (IABP) effects on patients' hemodynamics, some issues related to the effects of this therapy are still not fully understood. One of these issues is the effect of IABP, its inflation timing and duration on peripheral circulation autonomic controls. This work provides a systematic analysis of IABP effects on baroreflex using a cardiovascular hybrid model, which consists of computational and hydraulic submodels. The work also included a baroreflex computational model that was connected to a hydraulic model with a 40-cm(3) balloon. The IABP was operated at different inflation trigger timings (-0.14 to 0.31 s) and inflation durations (0.05-0.45 s), with time of the dicrotic notch being taken as t = 0. Baroreflex-dependent parameters-afferent and efferent pathway activity, heart rate, peripheral resistance, and venous tone-were evaluated at each of the inflation trigger times and durations considered. Balloon early inflation (0.09 s before the dicrotic notch) with inflation duration of 0.25 s generated a maximum net increment of afferent pathway activity of 10%, thus leading to a decrement of efferent sympathetic activity by 15.3% compared with baseline values. These times also resulted in a reduction in peripheral resistance and heart rate by 4 and 4.3% compared with baseline value. We conclude that optimum IABP triggering time results in positive effects on peripheral circulation autonomic controls. Conversely, if the balloon is not properly timed, peripheral resistance and heart rate may even increase, which could lead to detrimental outcomes.
© 2012, Copyright the Authors. Artificial Organs © 2012, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 23121229     DOI: 10.1111/j.1525-1594.2012.01540.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  3 in total

1.  Continuous-flow pump model study: the effect on pump performance of pump characteristics and cardiovascular conditions.

Authors:  Gianfranco Ferrari; Maciej Kozarski; Libera Fresiello; Arianna Di Molfetta; Krzysztof Zieliński; Krystyna Górczyńska; Krzysztof J Pałko; Marek Darowski
Journal:  J Artif Organs       Date:  2013-03-05       Impact factor: 1.731

2.  Framework for patient-specific simulation of hemodynamics in heart failure with counterpulsation support.

Authors:  Mattia Arduini; Jonathan Pham; Alison L Marsden; Ian Y Chen; Daniel B Ennis; Seraina A Dual
Journal:  Front Cardiovasc Med       Date:  2022-08-01

3.  A Model of the Cardiorespiratory Response to Aerobic Exercise in Healthy and Heart Failure Conditions.

Authors:  Libera Fresiello; Bart Meyns; Arianna Di Molfetta; Gianfranco Ferrari
Journal:  Front Physiol       Date:  2016-06-08       Impact factor: 4.566

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.