Development of a high-throughput electrophysiological assay for the human ether-à-go-go related potassium channel hERG.

Daniel J Gillie1, Steven J Novick, Brian T Donovan, Lisa A Payne, Claire Townsend.   

Abstract

INTRODUCTION: Drug-induced prolongation of the QT interval via block of the hERG potassium channel is a major cause of attrition in drug development. The advent of automated electrophysiology systems has enabled the detection of hERG block earlier in drug discovery. In this study, we have evaluated the suitability of a second generation automated patch clamp instrument, the IonWorks Barracuda, for the characterization of hERG biophysics and pharmacology.
METHODS: All experiments were conducted with cells stably expressing hERG. Recordings were made in perforated patch mode either on a conventional patch clamp setup or on the IonWorks Barracuda. On the latter, all recordings were population recordings in 384-well patch plates.
RESULTS: HERG channels activated with a V(1/2)=-3.2±1.6mV (n=178) on the IonWorks Barracuda versus -11.2±6.1mV (n=9) by manual patch clamp. On the IonWorks Barracuda, seal resistances and currents were stable (<30% change) with up to six cumulative drug additions and 1-min incubations per addition. Over 27 experiments, an average of 338 concentration-response curves were obtained per experiment (96% of the 352 test wells on each plate). HERG pharmacology was examined with a set of 353 compounds that included well-characterized hERG blockers. Astemizole, terfenadine and quinidine inhibited hERG currents with IC(50) values of 159nM, 224nM and 2μM, respectively (n=51, 10 and 18). This set of compounds was also tested on the PatchXpress automated electrophysiology system. We determined through statistical methods that the two automated systems provided equivalent results. DISCUSSION: Evaluating drug effects on hERG channels is best performed by electrophysiological methods. HERG activation and pharmacology on the IonWorks Barracuda automated electrophysiology platform were in good agreement with published electrophysiology results. Therefore, the IonWorks Barracuda provides an efficient way to study hERG biophysics and pharmacology.
Copyright © 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23103595     DOI: 10.1016/j.vascn.2012.10.002

Source DB:  PubMed          Journal:  J Pharmacol Toxicol Methods        ISSN: 1056-8719            Impact factor:   1.950


  13 in total

Review 1.  High throughput screening technologies for ion channels.

Authors:  Hai-bo Yu; Min Li; Wei-ping Wang; Xiao-liang Wang
Journal:  Acta Pharmacol Sin       Date:  2015-12-14       Impact factor: 6.150

Review 2.  Using automated patch clamp electrophysiology platforms in pain-related ion channel research: insights from industry and academia.

Authors:  Damian C Bell; Mark L Dallas
Journal:  Br J Pharmacol       Date:  2017-07-18       Impact factor: 8.739

3.  Compilation and physicochemical classification analysis of a diverse hERG inhibition database.

Authors:  Remigijus Didziapetris; Kiril Lanevskij
Journal:  J Comput Aided Mol Des       Date:  2016-10-25       Impact factor: 3.686

4.  Discovery and hit-to-lead evaluation of piperazine amides as selective, state-dependent NaV1.7 inhibitors.

Authors:  Brian A Sparling; S Yi; J Able; H Bregman; Erin F DiMauro; R S Foti; H Gao; A Guzman-Perez; H Huang; M Jarosh; T Kornecook; J Ligutti; B C Milgram; B D Moyer; B Youngblood; V L Yu; M M Weiss
Journal:  Medchemcomm       Date:  2016-12-02       Impact factor: 3.597

5.  Advancing Ion Channel Research with Automated Patch Clamp (APC) Electrophysiology Platforms.

Authors:  Damian C Bell; Mark L Dallas
Journal:  Adv Exp Med Biol       Date:  2021       Impact factor: 2.622

6.  Early identification of hERG liability in drug discovery programs by automated patch clamp.

Authors:  Timm Danker; Clemens Möller
Journal:  Front Pharmacol       Date:  2014-09-02       Impact factor: 5.810

7.  Variability in high-throughput ion-channel screening data and consequences for cardiac safety assessment.

Authors:  Ryan C Elkins; Mark R Davies; Stephen J Brough; David J Gavaghan; Yi Cui; Najah Abi-Gerges; Gary R Mirams
Journal:  J Pharmacol Toxicol Methods       Date:  2013-05-05       Impact factor: 1.950

8.  Prediction of Thorough QT study results using action potential simulations based on ion channel screens.

Authors:  Gary R Mirams; Mark R Davies; Stephen J Brough; Matthew H Bridgland-Taylor; Yi Cui; David J Gavaghan; Najah Abi-Gerges
Journal:  J Pharmacol Toxicol Methods       Date:  2014-07-31       Impact factor: 1.950

Review 9.  Concise Review: Measuring Physiological Responses of Human Pluripotent Stem Cell Derived Cardiomyocytes to Drugs and Disease.

Authors:  Berend J van Meer; Leon G J Tertoolen; Christine L Mummery
Journal:  Stem Cells       Date:  2016-06-30       Impact factor: 6.277

10.  Dynamic all-optical drug screening on cardiac voltage-gated ion channels.

Authors:  Jonas Streit; Sonja Kleinlogel
Journal:  Sci Rep       Date:  2018-01-18       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.