Suitability of POC lactate methods for fetal and perinatal lactate testing: considerations for accuracy, specificity and decision making criteria.

Jean-Luc Orsonneau1, François Fraissinet, Véronique Sébille-Rivain, Daniele Dudouet, Edith Bigot-Corbel.   

Abstract

BACKGROUND: Point of care (POC) testing for fetal scalp blood lactate is a more user friendly and more successful approach compared to scalp pH for intrapartum fetal monitoring. The aim of this study was to assess the analytical specificity and clinical reliability of three POC lactate methods.
METHODS: The analytical performance of three POC lactate methods was compared to Cobas 6000 (Roche Diagnostics) laboratory reference method: Lactate Pro from Arkray, GEM 4000 from Instrumentation Laboratory and StatStrip Lactate from Nova Biomedical. The clinical performance and influences on accuracy and decision making criteria for the three POC methods was assessed with umbilical cord samples and compared to the laboratory reference method. The influence of varying ranges of hemoglobin, pH and partial oxygen pressure (pO(2)) on the accuracy of results was assessed.
RESULTS: Although all three POC methods showed good correlation with the reference method for the umbilical cord sample population (r=0.989, 0.973 and 0.980, respectively), Lactate Pro and Gem 4000 showed a significant negative bias compared to the reference method. The degree of bias meant a significant readjustment of decision making criteria was required for fetal lactate use. The accuracy of the Lactate Pro results was affected by hemoglobin and to a lesser extent pH.
CONCLUSIONS: The three electrochemical POC devices can measure fetal lactate reliably. StatStrip Lactate showed a closer correlation and concordance to our laboratory reference method. The results of this study indicate the requirement for predetermining the reliability of POC lactate methods before use present in fetal and perinatal settings.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23096019     DOI: 10.1515/cclm-2012-0201

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

1.  Point of Care Testing of Serum Electrolytes and Lactate in Sick Children.

Authors:  Aashima Dabas; Shipra Agrawal; Vernika Tyagi; Shikha Sharma; Vandana Rastogi; Urmila Jhamb; Pradeep Kumar Dabla
Journal:  EJIFCC       Date:  2021-06-29

2.  A Point-of-Care Serum Lactate Level and Mortality in Adult Sepsis Patients: A Community Hospital Setting.

Authors:  Suraphan Charoentanyarak; Bundit Sawunyavisuth; Sansanee Deepai; Kittisak Sawanyawisuth
Journal:  J Prim Care Community Health       Date:  2021 Jan-Dec

3.  Protocol for a randomised controlled trial of fetal scalp blood lactate measurement to reduce caesarean sections during labour: the Flamingo trial [ACTRN12611000172909].

Authors:  Christine E East; Stefan C Kane; Mary-Ann Davey; C Omar Kamlin; Shaun P Brennecke
Journal:  BMC Pregnancy Childbirth       Date:  2015-11-03       Impact factor: 3.007

4.  Lactate point-of-care testing for acidosis: Cross-comparison of two devices with routine laboratory results.

Authors:  Remco van Horssen; Teska N Schuurman; Monique J M de Groot; Bernadette S Jakobs
Journal:  Pract Lab Med       Date:  2015-12-24
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.