Literature DB >> 23089834

[Comparison of various image guided radiation therapy systems; image-guided localization accuracy and patient throughput].

Eriko Takenaga1, Yuji Nakaguchi, Masato Maruyama, Nozomu Nagasue, Kiyotaka Kakei, Yudai Kai, Tomohiro Kouno, Motoharu Sasaki, Masahiro Hashida.   

Abstract

In this study, we evaluated various image guided radiation therapy (IGRT) systems regarding accuracy and patient throughput for conventional radiation therapy. We compared between 2D-2D match (the collation by 2 X-rays directions), cone beam computed tomography (CBCT), and ExacTrac X-Ray system using phantom for CLINAC iX and Synergy. All systems were able to correct within almost 1 mm. ExacTrac X-Ray system showed in particular a high accuracy. As for patient throughput, ExacTrac X-Ray system was the fastest system and 2D-2D match for Synergy was the slowest. All systems have enough ability with regard to accuracy and patient throughput on clinical use. ExacTrac X-Ray system showed superiority with accuracy and throughput, but it is important to note that we have to choose the IGRT technique depending on the treatment site, the purpose, and the patient's state.

Entities:  

Mesh:

Year:  2012        PMID: 23089834     DOI: 10.6009/jjrt.2012_jsrt_68.10.1327

Source DB:  PubMed          Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi        ISSN: 0369-4305


  1 in total

1.  Model for Estimating Power and Downtime Effects on Teletherapy Units in Low-Resource Settings.

Authors:  Rachel McCarroll; Bassem Youssef; Beth Beadle; Maureen Bojador; Rex Cardan; Robin Famiglietti; David Followill; Geoffrey Ibbott; Anuja Jhingran; Christoph Trauernicht; Peter Balter; Laurence Court
Journal:  J Glob Oncol       Date:  2017-01-11
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.