Potential sources of interference on Abeta immunoassays in biological samples.

Hugo Vanderstichele1, Erik Stoops1, Eugeen Vanmechelen1, Andreas Jeromin2.   

Abstract

Therapeutic products that depend on the use of an in vitro diagnostic biomarker test to confirm their effectiveness are increasingly being developed. Use of biomarkers is particularly meaningful in the context of selecting the patient population where the therapeutic treatment is believed to be efficacious (patient enrichment). Currently available 'research-use-only' assays for Alzheimer's disease diagnosis all suffer from non-analyte and analyte-specific interferences. The impact of these interferences on the outcome of the assays is not well understood. The confounding factors are hampering correct value determination in biological samples and are intrinsic to the assay concept, the assay design, the presence in the sample of heterophilic antibodies and auto-antibodies, or might be the result of the therapeutic approach. This review focuses on the importance of assay interferences and considers how these might be minimized with the final aim of making the assays more acceptable as in vitro diagnostic biomarker tests for theranostic use.

Entities:  

Year:  2012        PMID: 23082750      PMCID: PMC3580396          DOI: 10.1186/alzrt142

Source DB:  PubMed          Journal:  Alzheimers Res Ther            Impact factor:   6.982


Introduction

The development of therapeutic compounds that depend on the use of an in vitro diagnostic biomarker test (IVD) to confirm their effectiveness will become more common in the future. Companion diagnostics will ultimately shorten the development time for Alzheimer's disease (AD) therapeutic trials and increase their success rates. When the therapeutic product becomes available, assay information will be used to select (stratification) or exclude (risk assessment) patient populations for a particular clinical study, to optimize dosing regimens, or to identify subjects who will most likely respond to treatment and will not suffer from side effects (responders, safety). If the outcome of a diagnostic assay determines how a patient will be treated, it is obvious that health care professionals must be able to rely on the quality of the result. Inadequate performance characteristics of an IVD or companion diagnostic biomarker test could expose a patient to preventable treatment risks. Several research assays for AD biomarkers in cerebrospinal fluid (CSF) evolved over the past decade from proof-of-concept to tools with promising or accepted clinical value. In this disease field, no US Food and Drug Administration-approved assay is available yet on the market, due in part to some drawbacks in their analytical performance characteristics. The US Food and Drug Administration provides more detailed relevant policies for the safety and effectiveness of IVD companion diagnostic devices as used with therapeutics [1]. The AD community has considered for several decades that the β-amyloid protein (Aβ) might be at the origin of AD, although amyloidopathy is not absolutely specific for AD [2-4]. A full understanding of its clinical relevance is hampered by (i) the intrinsic nature of Aβ, including its aggregation and adsorption properties, (ii) the complexity and heterogeneity of Aβ isoforms, including modifications or different conformational forms, (iii) the presence of confounding factors, (iv) low concentrations of Aβ in biological fluids, (v) high variability in outcomes of each assay between study centers, and (vi) the absence of a reference method or reference materials (relative quantitative assays) [5,6].

Problem statement

Immunoassays that use antibodies are easy to perform, specific for an epitope or conformation of an analyte, and highly vulnerable towards confounding factors or interferences [5] (in this context, an interference is an effect of a substance present in the sample that alters the correct value of the result). Detailed understanding of the nature, the prevalence, the complexity, the technology- or protocol-dependency, as well as the interactions between different confounding factors is key to define solutions and improve the robustness of the test methods. Cost-efficient and user-friendly integration in the product design of assay modifications to reduce interferences, without having an impact on the clinical accuracy, is a major challenge. Assay interferences are often underestimated, but highly relevant; they have an effect on sample homogeneity and stability, assay precision, or clinical interpretation. Every false result will generate extra cost for the lab and will introduce preventable concerns (through the incorrect message given) for patients, families, and caregivers. Immunoassays measure the presence (qualitative assay), concentrations (quantitative assay), or changes in concentrations of one or several analytes in a complex mixture of proteins. The affinity of the antibody for the analyte is related to its thermodynamic property (association and dissociation capacity). Antibodies and antigens (or antigen conformations) are in a state of dynamic equilibrium that is concentration dependent. Only a fraction of the total amount of analyte might be detectable by the immunoassays. Notwithstanding the well-known pre-analytical variables [5], the measurement of Aβ by classical immunoassays is complicated by artificial or induced confounding factors, which are illustrated in Figure 1 and discussed here. This review will not focus on antibody-independent techniques, as this could be the subject of future discussions, but discusses in more detail the confounding factors and some possibilities for overcoming them.
Figure 1

Interferences observed in assays for quantification of β-amyloid. The figure provides a summary on how endogenous antibodies can interfere in immunoassays measuring β-amyloid (Aβ). The box visualizes the complexity (i) between bound and unbound analyte, or (ii) between monomeric and aggregated analyte.

Interferences observed in assays for quantification of β-amyloid. The figure provides a summary on how endogenous antibodies can interfere in immunoassays measuring β-amyloid (Aβ). The box visualizes the complexity (i) between bound and unbound analyte, or (ii) between monomeric and aggregated analyte.

Non-analyte-specific interference

Confounding factors

Non-analyte-specific interferences are not necessarily directly linked to one specific analyte, but might be relevant also for other proteins in the sample. Several non-analyte-specific parameters in the product design have a direct effect on the equilibrium constant of the antigen-antibody reaction (for example, temperature, pH, ionic strength), while others have not (for example, antigen and antibody concentration, duration of incubation) [7]. Underdeveloped parameters can explain discrepancies between study results. Problems occur at the level of the assay, the raw materials, or the sample. For example, the pH of the assay diluent modifies the analytical sensitivity of Aβ1-42 assays [8] and affects the oligomerization state of Aβ [9]. It is well-known that the pH of CSF increases rapidly after storage at room temperature for a few hours [10], although Bjerke and colleagues [10] did not observe any difference in Aβ1-42 concentration when CSF samples were tested at different pH levels. Aβ1-42 detection in CSF is further influenced by temperature, especially within the AD group [11], affecting its diagnostic use. Non-specific binding can occur also when other components (for example, secondary conjugates, detector mAbs, substrate) directly bind to the solid phase or when aggregated proteins in a β-pleated sheet conformation bind to each other. The latter was documented for biotinylated mAbs depending on the pH of the storage buffer [12], an effect that can be prevented by the use of arginine [13]. Each component in the product has to be screened for its influence on the intended use of the assay. If needed, test instructions have to be re-evaluated, even when the product is already available on the market.

Heterophilic antibodies

Heterophilic antibodies (HAs) are found in a number of healthy and diseased patient samples (5 to 40%) [14]. These endogenous (mostly polyclonal human) antibodies have a broad spectrum of activity against non-immune immunoglobulins from several species, as well as reactivity towards poorly defined (self) antigens. In contrast, human anti-animal antibodies are directed against well-defined antigens and show high affinity. Their presence can be the consequence of the administration of an exogenous antibody, such as from treatment with therapeutic antibodies. The relevance, occurrence, and importance of HAs as a confounding factor for the measurement of CSF Aβ have yet to be fully described. HAs have weak affinity and are multispecies specific. The interfering antibodies are of the IgG, IgM or IgA type. Generally, HAs are directed to the Fc part (anti-isotypical interference, such as rheumatoid factor), while antiidiotypical interfering antibodies bind to the highly variable Fab portion of the molecule (an antiidiotypical antibody is directed towards the antigen-binding site of another antibody; the antigen binding site of this antibody can be similar in structure to the original antigen). The concentration of HAs is higher in blood samples than in CSF [15]. If plasma samples from the same subject are collected and tested over time, the interference will be visible in each sample from the affected subject, independent of the technology platform. Also, for low abundant proteins such as Aβ oligomers, positive signals are eliminated (plasma) or reduced (CSF) when assays are repeated in the presence of HA-blocking factors [16]. Small amounts of HAs will immediately affect the quantification of low-abundance proteins, especially when the dilution factor for the sample in the assay is limited, unless HA-reducing buffers were included in the test concept during the development phase of the product. HAs are not directly visible in single analyte immunoassays. They can be identified by replacement of one antibody (capture mAb, detector mAb) with a non-analyte-specific antibody [17], using antibodies from another species, by selection of assays from other vendors, or (maybe in part) by the absence of parallelism in serial dilutions of the sample. In the latter case, no experimental evidence is available to date that HAs are the major confounders. HA interference can be minimized by assay modifications or sample pre-treatment. At the assay level, interference is minimized by (i) replacement of mAbs by Fab fragments, humanized antibodies, affibodies or aptamers [18], (ii) the use of ready-to-use blocking buffers, detergentia, heterophilic blocking tubes (for example, Scantibodies) or affinity discriminator buffers (for example, LowCross buffer from CANDOR Biosciences, Wangen im Allgäu, Germany), or (iii) the addition of well-con-trolled non-immune mouse IgG or animal serum (this will not neutralize the anti-idiotypic or anti-anti-idiotypic forms of HAs [19]). Sample-pretreatment procedures have already been implemented in commercial assays for Aβ1-42. Increased levels of free Aβ1-42 were noted after dilution of EDTA-plasma samples in detergent-containing buffer [20] or after dilution of CSF [10,21,22]. If compatible with the properties of the analyte, HAs can be removed by heating [8,10], although this is not a customer-friendly approach and might be detrimental to non-Aβ analytes to be analyzed in the sample. Bjerke and colleagues [10] reported a higher increase of CSF Aβ1-42 after heating of AD samples, resulting in reduced clinical accuracy. Extraction of HAs prior to analysis by depletion of IgG (protein A, protein G) [16], polyethyleneglycol precipitation, or pre-absorption with human gamma globulin-coated beads/Protein L-coated beads [23] has been proposed for other analytes. These extraction procedures will add more cost and workload to the assay used in routine clinical testing labs. The improvement methods described above might be an excellent approach for one analyte, detrimental to another, and only applicable to one specific sample type. The exact impact of any assay changes on clinical outcome has to be qualified.

Analyte-specific interference

Aβ autoantibodies

Analyte-specific interferences will directly compete with the quantification of Aβ in immunoassays. The prevalence of analyte-specific interferences in modern (blocked) two-site immunoassays is very low (< 0.05%). The relevance, occurrence, and importance of auto-antibodies as a confounding factor for CSF Aβ measurements are uncertain at present due to a lack of well-characterized and validated detection systems and are a subject for future studies. The clinical relevance of auto-antibodies against Aβ is still unclear, related in part to the urgent need for harmonization of the detection methodologies. The introduction of Aβ immunotherapies will further accelerate the development of harmonized immune biomarkers (Table 1).
Table 1

Passive immunotherapy with monoclonal antibodies

Drug nameSponsorStatusAβ epitopeReference
Passive immunotherapy with mAbs
 BapineuzumabJ & JIII1-5[36,37]
ElanConformational and linear
Pfizer
 SolanezumabLillyIII13-28[38,39]
Linear
 GantenerumabRocheIIConformational[40,41]
 PonezumabPfizerII (stopped)33-40[42]
Linear
 CrenezumabGenentechII1-15[43]
Conformational
 BAN2401EisaiIAmino-terminal
Conformational
Passive immunotherapy with intravenous IgGs
 OctagamOctapharmaII[44]
 NewgamSutter HealthII
Active immunotherapy, vaccines
 ACC-001J & JII1-7[45]
Wyeth
 CAD106NovartisII1-6[46]
 ACI-24AC ImmuneII1-15[43]
 Affitope AD02GSKAffirisIIMimick of amino terminus[47]

Aβ, β-amyloid; GSK, GlaxoSmithKline; J & J, Johnson and Johnson.

Passive immunotherapy with monoclonal antibodies Aβ, β-amyloid; GSK, GlaxoSmithKline; J & J, Johnson and Johnson. Auto-antibodies in AD can influence the IgG-based clearance system for amyloidogenic proteins [24], modulate plaque removal [25], be indicative for dysfunctional immune signaling [26], reflect blood-brain barrier break-down [27], or reduce neuritic dystrophy and astrogliosis in mouse models [28]. When low concentrations of an analyte under investigation are present in a sample, such as in the case of Aβ oligomers as a potential neurotoxin, auto-antibodies may pose additional particular challenges linked to the analytical sensitivity of the technology platforms used for their detection [16]. Naturally occurring antibodies against Aβ are found in CSF and plasma of healthy and diseased subjects. Patients with AD have lower, higher, or identical levels of serum anti-Aβ antibodies to healthy age-matched individuals [24]. The serum anti-Aβ antibody titer was higher in patients with mild cognitive impairment who progressed to AD than in stable cases [29]. The level of auto-antibodies to Aβ did not correlate with the 'likeli-hood of developing AD [30], although titers were negatively correlated with the cognitive status [31]. Henkel and colleagues [24] showed that the concentration of the large Aβ-binding particles (LAPs) was highly variable among individuals and that the levels were not disease-specific. The abovementioned studies use different technologies and are not always qualified extensively for the intended use. The immunoreactivity was 30 to 230 times lower in CSF than in plasma [32]. Anti-Aβ IgGs were restricted to the IgG1 and IgG3 subclass, which are the most potent IgG subclasses in activating the complement classical pathway through C1q binding [33]. Several types of antibodies recognize the higher order Aβ assemblies, including oligomers and post-translationally modified peptides, rather than the monomeric forms [32]. The signal in unvaccinated nonTg mice was specific for the full-length Aβ peptide, while for anti-Aβ vaccines, the pre-dominant epitope is in the amino-terminal domain [34]. In the study of Henkel and colleagues [24], confocal microscopy for single molecular imaging was used to identify and classify distinct types of LAPs. Double labeling with anti-Aβ and anti-IgG antibodies characterized LAP-3 and LAP-4 as immune complexes of Aβ and IgG. The LAPs are most probably directed against either a non-pathogenic form of Aβ or aggregated Aβ1-42. The antibody-antigen complex can be dissociated by incubation at low pH [34,35]. The increase in Aβ auto-antibody levels is higher in AD than in controls, again possibly affecting clinical accuracy. Differences between studies may relate to a diversity of methods used for anti-Aβ assessment, the validation status of the assays, and detailed investigation of the relevance of the generated output. One approach to solve this issue could be the development of a multi-analyte assay format, either (i) by inclusion of an extra region with a non-analyte specific mAb used as capture antibody (direct bridging between mAbs) [17], or (ii) by the integration of an internal control with Aβ coupled to a specific region to verify the presence of anti-Aβ antibodies. This approach will help to understand the contribution of Aβ auto-antibodies to the output signal of the assay.

Analyte-specific interferences induced by therapeutics

Many pharmaceutical and biotechnology companies perform studies with Aβ-lowering drugs: vaccination with peptides (active) or immunization with mAbs (passive immunization) (Table 1). Vaccination improves the brain pathology and protects against cognitive decline, while passive immunization with Aβ antibodies reduces plaque burden [2]. Therapeutic compounds, however, can directly interfere with assay design and assay modifications are required dependent on the therapeutic approach. Short amino-terminal-specific Aβ-peptides used for immunization will compete for binding with the mAbs if amino-terminal-specific mAbs are used, while the therapeutic mAbs can mask the binding site of the analyte, resulting in an underestimation of the levels in a sample. Analyte-specific interferences induced by a specific therapeutic (with a defined molecular nature) can be identified easily and reproducibly by spiking of the drug in relevant concentrations in the sample of interest, followed by its quantification using different technologies and/or antibody pair combinations. In active immunization trials with a more heterogeneous antibody spectrum, it might be more relevant to include antibody-independent techniques (for example, mass spectrometry). Although the latter might not be relevant for verification of efficacy of drug treatment since these techniques quantify the total amount of Aβ in a sample. The monomeric, non-bound Aβ could be the most relevant Aβ isoform to verify for a disease-modifying effect. As such, it will be obligatory to develop assays in biological fluids that make a distinction between free analyte, and the analyte bound to confounding factors (for example, endogenous antibodies, therapeutic compounds), or conformational forms. Comparison with the total amount of Aβ, independent of the isoform, could be an extra advantage for clinical interpretation of the data.

Conclusion

Measurements of low concentrations of Aβ1-42 in biological samples are impeded by the presence of non-specific interactions. These documented confounders need to be addressed during the development phase of the product to improve robustness, respecting the balance between the required assay sensitivity and precision. The impact of assay modifications on the accuracy of clinical decisions still has to be determined.

Abbreviations

Aβ: β-amyloid; AD: Alzheimer's disease; CSF: cerebrospinal fluid; HA: heterophilic antibodies; IVD: in vitro diagnostic biomarker test; LAP: large Aβ-binding particle; mAb: monoclonal antibody.

Competing interests

The authors declare that they have no competing interests.
  45 in total

1.  Elevation of {beta}-amyloid 1-42 autoantibodies in the blood of amnestic patients with mild cognitive impairment.

Authors:  Daniela Storace; Sergio Cammarata; Roberta Borghi; Roberta Sanguineti; Luca Giliberto; Alessandra Piccini; Valeria Pollero; Cristina Novello; Carlo Caltagirone; Mark A Smith; Paola Bossù; George Perry; Patrizio Odetti; Massimo Tabaton
Journal:  Arch Neurol       Date:  2010-07

2.  Improving immunoassay performance by antibody engineering.

Authors:  Ulf-Håkan Stenman
Journal:  Clin Chem       Date:  2005-05       Impact factor: 8.327

3.  Improved multiplex immunoassay performance in human plasma and synovial fluid following removal of interfering heterophilic antibodies.

Authors:  Wilco de Jager; Berent J Prakken; Johannes W J Bijlsma; Wietse Kuis; Ger T Rijkers
Journal:  J Immunol Methods       Date:  2005-05       Impact factor: 2.303

4.  Brain-reactive autoantibodies prevalent in human sera increase intraneuronal amyloid-β(1-42) deposition.

Authors:  Robert G Nagele; Peter M Clifford; Gilbert Siu; Eli C Levin; Nimish K Acharya; Min Han; Mary C Kosciuk; Venkat Venkataraman; Semah Zavareh; Shabnam Zarrabi; Kristin Kinsler; Nikhil G Thaker; Eric P Nagele; Jacqueline Dash; Hoau Y Wang; Andrew Levitas
Journal:  J Alzheimers Dis       Date:  2011       Impact factor: 4.472

5.  Antigen-antibody dissociation in Alzheimer disease: a novel approach to diagnosis.

Authors:  Katarzyna A Gustaw; Matthew R Garrett; Hyoung-Gon Lee; Rudy J Castellani; Michael G Zagorski; Annamalai Prakasam; Sandra L Siedlak; Xiongwei Zhu; George Perry; Robert B Petersen; Robert P Friedland; Mark A Smith
Journal:  J Neurochem       Date:  2008-05-12       Impact factor: 5.372

6.  Liposomal vaccines with conformation-specific amyloid peptide antigens define immune response and efficacy in APP transgenic mice.

Authors:  Andreas Muhs; David T Hickman; Maria Pihlgren; Nathalie Chuard; Valérie Giriens; Carine Meerschman; Ingrid van der Auwera; Fred van Leuven; Masae Sugawara; Marie-Catherine Weingertner; Burkhard Bechinger; Ruth Greferath; Nadine Kolonko; Luitgard Nagel-Steger; Detlev Riesner; Roscoe O Brady; Andrea Pfeifer; Claude Nicolau
Journal:  Proc Natl Acad Sci U S A       Date:  2007-05-21       Impact factor: 11.205

7.  Patients with Alzheimer disease have lower levels of serum anti-amyloid peptide antibodies than healthy elderly individuals.

Authors:  Marc E Weksler; Norman Relkin; Rimma Turkenich; Susan LaRusse; Ling Zhou; Paul Szabo
Journal:  Exp Gerontol       Date:  2002-07       Impact factor: 4.032

8.  Neuroprotective natural antibodies to assemblies of amyloidogenic peptides decrease with normal aging and advancing Alzheimer's disease.

Authors:  M Britschgi; C E Olin; H T Johns; Y Takeda-Uchimura; M C LeMieux; K Rufibach; J Rajadas; H Zhang; B Tomooka; W H Robinson; C M Clark; A M Fagan; D R Galasko; D M Holtzman; M Jutel; J A Kaye; C A Lemere; J Leszek; G Li; E R Peskind; J F Quinn; J A Yesavage; J A Ghiso; T Wyss-Coray
Journal:  Proc Natl Acad Sci U S A       Date:  2009-07-06       Impact factor: 11.205

9.  Improvement of a low pH antigen-antibody dissociation procedure for ELISA measurement of circulating anti-Abeta antibodies.

Authors:  Qingyou Li; Marcia Gordon; Chuanhai Cao; Kenneth E Ugen; Dave Morgan
Journal:  BMC Neurosci       Date:  2007-03-20       Impact factor: 3.288

10.  Inhibition of protein aggregation: supramolecular assemblies of arginine hold the key.

Authors:  Utpal Das; Gururao Hariprasad; Abdul S Ethayathulla; Pallavi Manral; Taposh K Das; Santosh Pasha; Anita Mann; Munia Ganguli; Amit K Verma; Rajiv Bhat; Sanjeev Kumar Chandrayan; Shubbir Ahmed; Sujata Sharma; Punit Kaur; Tej P Singh; Alagiri Srinivasan
Journal:  PLoS One       Date:  2007-11-14       Impact factor: 3.240

View more
  3 in total

1.  Quantitative detection of α-Synuclein and Tau oligomers and other aggregates by digital single particle counting.

Authors:  Lara Blömeke; Marlene Pils; Victoria Kraemer-Schulien; Alexandra Dybala; Anja Schaffrath; Andreas Kulawik; Fabian Rehn; Anneliese Cousin; Volker Nischwitz; Johannes Willbold; Rebecca Zack; Thomas F Tropea; Tuyen Bujnicki; Gültekin Tamgüney; Daniel Weintraub; David Irwin; Murray Grossman; David A Wolk; John Q Trojanowski; Oliver Bannach; Alice Chen-Plotkin; Dieter Willbold
Journal:  NPJ Parkinsons Dis       Date:  2022-06-02

2.  The Effect of Propofol Versus Isoflurane Anesthesia on Human Cerebrospinal Fluid Markers of Alzheimer's Disease: Results of a Randomized Trial.

Authors:  Miles Berger; Jacob W Nadler; Allan Friedman; David L McDonagh; Ellen R Bennett; Mary Cooter; Wenjing Qi; Daniel T Laskowitz; Vikram Ponnusamy; Mark F Newman; Leslie M Shaw; David S Warner; Joseph P Mathew; Michael L James
Journal:  J Alzheimers Dis       Date:  2016-04-15       Impact factor: 4.472

Review 3.  Amyloid Beta and Tau as Alzheimer's Disease Blood Biomarkers: Promise From New Technologies.

Authors:  Lih-Fen Lue; Andre Guerra; Douglas G Walker
Journal:  Neurol Ther       Date:  2017-07-21
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.