Laboratory analysis of phacoemulsifier compliance and capacity.

Mitra Nejad1, Valentine P Injev, Kevin M Miller.   

Abstract

PURPOSE: To compare the compliance and capacity of 7 fluidics modules used by 6 phacoemulsifiers from 3 manufacturers.
SETTING: Jules Stein Eye Institute, Los Angeles, California, USA.
DESIGN: Experimental study.
METHODS: Previous-model and current-model phacoemulsifiers from 3 manufacturers were subjected to laboratory analysis of compliance and capacity. Previous-generation models tested included the Legacy Advantec, Whitestar Sovereign Phacoemulsification System, and Millennium Microsurgical System. Current models tested were the Infiniti Vision System with standard and Intrepid cassettes, Whitestar Signature Phacoemulsification System, and Stellaris Vision Enhancement System. To measure compliance, the aspiration line was connected to an electronic pressure transducer and small volumes of fluid were injected or aspirated. To measure capacity, the space between the distal end of the aspiration line and the pump was filled with methylene blue-dyed fluid.
RESULTS: The Legacy was the most compliant phacoemulsifier. The old and new Whitestar systems, Millennium system, and Stellaris system showed similar midrange compliances. The Infiniti Vision System with the Intrepid fluidic management system was the least compliant. The Infiniti cassettes had the greatest capacity, which is a detriment from a surge-control perspective, and Signature cassettes had the least capacity.
CONCLUSIONS: The Infiniti Intrepid system had the lowest compliance of the 6 units tested, which is optimum from a surge-control perspective. All other things being equal, the Infiniti should have the safest occlusion-break surge response. FINANCIAL DISCLOSURE: Mr. Injev is an employee of Alcon Laboratories. Dr. Miller is a consultant to and investigator for Alcon Laboratories. Ms. Nejad has no financial or proprietary interest in any material or method mentioned.
Copyright © 2012 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23079316     DOI: 10.1016/j.jcrs.2012.05.046

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Elevated intraocular pressure causes cellular and molecular retinal injuries, advocating a more moderate intraocular pressure setting during phacoemulsification surgery.

Authors:  Zhenni Zhao; Xiaowei Yu; Xue Yang; Jiamin Zhang; Dandan Zhang; Nannan Sun; Zhigang Fan
Journal:  Int Ophthalmol       Date:  2020-07-28       Impact factor: 2.031

2.  Comparison of occlusion break responses and vacuum rise times of phacoemulsification systems.

Authors:  Pooria Sharif-Kashani; Douglas Fanney; Val Injev
Journal:  BMC Ophthalmol       Date:  2014-07-30       Impact factor: 2.209

3.  Active-fluidics versus gravity-fluidics system in phacoemulsification for age-related cataract (AGSPC): study protocol for a prospective, randomised, double-blind, controlled clinical trial.

Authors:  Yu Luo; Hongyu Li; Wenqian Chen; Yi Gao; Tianju Ma; Zi Ye; Zhaohui Li
Journal:  BMJ Open       Date:  2022-01-20       Impact factor: 2.692

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.