Literature DB >> 23078925

Description of a new, endoscopic technique to remove the over-the-scope-clip in an ex vivo porcine model (with video).

Helmut Neumann1, Hiwot Diebel, Klaus Mönkemüller, Andreas Nägel, Dane Wildner, Michael Vieth, Jürgen Siebler, Markus F Neurath.   

Abstract

BACKGROUND: Various studies have demonstrated the usefulness of the over-the-scope-clip (OTSC) to treat perforations, anastomotic leaks, and fistulae. Endoscopic removal of the OTSC was previously described in a series of 3 patients by using the Nd:YAG laser.
OBJECTIVE: To evaluate a new endoscopic technique to remove the OTSC.
DESIGN: Prospective, single-arm, pilot study in an ex vivo porcine model.
INTERVENTIONS: Perforations were created by using a surgical scalpel and a blunt trocar. Then they were endoscopically closed with the OTSC. Next, the OTSC was removed under endoscopic control by inserting a 0.035-inch straight hydrophilic tip guidewire into the oval hole on the side of the OTSC jaw. Afterward, the OTSC clip was removed by pulling on the wire.
RESULTS: Fifteen perforations were closed: with the OTSC. In all of the cases, the endoscopic closure of the defects was feasible and effective. Successful visualization of the oval hole of the OTSC was possible in 12 cases (80%), and guidewire cannulation was possible in all of these 12 cases (100%). Advancement of the guidewire through the OTSC and then the lumen of the stomach was accomplished in 8 cases (53.3%). In all of the cases with successful cannulation of the orifice, removal of the OTSC was managed safely. The result was an overall success rate of 53.3% (8 of 15 cases). LIMITATIONS: Ex vivo porcine model.
CONCLUSIONS: Guidewire removal is a new and feasible technique to remove the OTSC. Future studies should refine the technique to enhance visualization and cannulation of the oval hole of the OTSC.
Copyright © 2012 American Society for Gastrointestinal Endoscopy. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23078925     DOI: 10.1016/j.gie.2012.07.036

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  7 in total

1.  DC current pulses for OTSC clip fragmentation: technology and experimental study.

Authors:  Sebastian Schostek; C N Ho; M Melbert; M Krautwald; A Caputo; G Parisi; M Wehrmann; T Gottwald; M O Schurr
Journal:  Surg Endosc       Date:  2014-11-01       Impact factor: 4.584

2.  Endoscopic removal of over-the-scope clips: Clinical experience with a bipolar cutting device.

Authors:  Markus Bauder; Benjamin Meier; Karel Caca; Arthur Schmidt
Journal:  United European Gastroenterol J       Date:  2016-10-06       Impact factor: 4.623

3.  Endoscopic closure of a gastrocolic fistula using the over-the-scope-clip-system.

Authors:  Klaus Mönkemüller; Shajan Peter; Basem Alkurdi; Jayapal Ramesh; Daniel Popa; C Mel Wilcox
Journal:  World J Gastrointest Endosc       Date:  2013-08-16

4.  One hundred and one over-the-scope-clip applications for severe gastrointestinal bleeding, leaks and fistulas.

Authors:  Edris Wedi; Susana Gonzalez; Detlev Menke; Elena Kruse; Kai Matthes; Juergen Hochberger
Journal:  World J Gastroenterol       Date:  2016-02-07       Impact factor: 5.742

5.  Over-the-scope clip (OTSC) reduces surgery rate in the management of iatrogenic gastrointestinal perforations.

Authors:  Sherine Khater; Gabriel Rahmi; Guillaume Perrod; Elia Samaha; Hedi Benosman; Leila Abbes; Georgia Malamut; Christophe Cellier
Journal:  Endosc Int Open       Date:  2017-05

Review 6.  Evolving techniques for gastrointestinal endoscopic hemostasis treatment.

Authors:  Kevin A Ghassemi; Dennis M Jensen
Journal:  Expert Rev Gastroenterol Hepatol       Date:  2016-01-11       Impact factor: 3.869

Review 7.  Current innovations in endoscopic therapy for the management of colorectal cancer: from endoscopic submucosal dissection to endoscopic full-thickness resection.

Authors:  Shintaro Fujihara; Hirohito Mori; Hideki Kobara; Noriko Nishiyama; Tae Matsunaga; Maki Ayaki; Tatsuo Yachida; Asahiro Morishita; Kunihiko Izuishi; Tsutomu Masaki
Journal:  Biomed Res Int       Date:  2014-04-30       Impact factor: 3.411

  7 in total