Oyinlolu O Adeyanju, Haitham M Al-Angari, Alan V Sahakian.
Abstract
BACKGROUND: Irreversible electroporation (IRE) is a novel ablation tool that uses brief high-voltage pulses to treat cancer. The efficacy of the therapy depends upon the distribution of the electric field, which in turn depends upon the configuration of electrodes used.Entities:
Keywords: hepatocellular carcinoma; irreversible electroporation; optimization, electrode configuration
Year: 2012 PMID: 23077449 PMCID: PMC3472940 DOI: 10.2478/v10019-012-0026-y
Source DB: PubMed Journal: Radiol Oncol ISSN: 1318-2099 Impact factor: 2.991
FIGURE 1Depiction of the electrode configurations. (left) The purple circles represent active electrodes, and the lighter shaded electrodes are set to ground. The distance (d1) between electrodes and the depth of insertion (d2) was varied in the study.
FIGURE 2A–C. Graphical depiction of VIRE values upon distance at several depths. There appears to be a parabolic minima at 2 cm distance between the electrodes for the two electrode set, whereas the VIRE values for both three and four electrodes decrease linearly with the distance between the electrodes. D–F Graphical depiction of the dependence of the VANT values upon distance at several depths. The VANT values decrease with distance for all electrode setups. Two electrodes overall have the lowest VANT values, followed by four electrodes and then three electrodes.
FIGURE 3Cross section of the tumor (circle) and its ablation zone (blue) at an applied voltage of 5.05 kV, which corresponds to the VIRE for two electrodes at a distance of 2.5 cm and depth of 1 cm for A) two electrodes B) three electrodes C) four electrodes D) three electrodes with the center of the electrodes shifted to the right 1 cm Note the asymmetry in B and the narrowing of the ablation zone in C. A bounding box around the tumor was used in the FEM simulations to improve the quality of the meshing and computations.
Summary statistics from the three VIRE tables (V)
| Two | 3700 | 5296 | 8500 |
| Three | 7500 | 10761 | 18700 |
| Four | 4700 | 7279 | 13600 |
|
| |||
|
| |||
| Parameter varying | Two electrodes | Three electrodes | Four electrodes |
| Distance | 6343 | 16036 | 12873 |
| Depth | 2133 | 1616 | 1762 |
Summary statistics from the three VANT tables (cm3)
| Two | 3.53 | 12.1 | 35.1 |
| Three | 13.2 | 60.5 | 171.3 |
| Four | 11.5 | 60.7 | 178.6 |
|
| |||
|
| |||
| Parameter varying | Two electrodes | Three electrodes | Four electrodes |
| Distance | 41.2 | 221 | 236 |
| Depth | 11.4 | 12.5 | 20.0 |