Literature DB >> 23073061

Outflow failure caused by mesothelial cell lining sheet wrapping in a patient with peritoneal dialysis.

Seok Hui Kang, Dong Shik Lee, Jong Won Park.   

Abstract

A 60-year-old male was admitted to our hospital due to nausea and poor oral intake. He had received distal pancreatectomy due to a traffic accident 15 years ago, and therefore, had a midline abdominal scar. Peritoneal dialysis (PD) was selected for renal replacement therapy. PD catheter was inserted by a surgical method with left paramedian incision under local anesthesia. Initially, dialysate flow was good, but outflow failure developed on postoperative day nine. Even after the use of laxatives and a fibrinolytic agent to correct the catheter obstruction, the problem was not resolved. Therefore, we performed an open laparotomy. When PD catheter was examined during laparotomy, a single thick band (0.6 cm × 36 cm), without blood vessels was wrapped around it. One of the ends of this band originated from the jejunum and the other was attached on the most proximal side hole of the catheter. A band was removed surgically, and then the PD catheter was repositioned in the abdominal cavity. On post-revision Day 7, PD started again and dialysate flow was good. The pathologic finding was consistent with mesothelial cell lining sheet. Considering its location and the history of a major abdominal operation, the adhesive band or sheet may be the cause in this case.

Entities:  

Mesh:

Year:  2014        PMID: 23073061     DOI: 10.5414/CN107562

Source DB:  PubMed          Journal:  Clin Nephrol        ISSN: 0301-0430            Impact factor:   0.975


  2 in total

1.  Malfunction of Peritoneal Catheters by Proliferation of Vascularized Fibrous Tissue.

Authors:  Matthias Zeiler; Federica F Lenci; Rosa M Agostinelli; Tania Monteburini; Rita Marinelli; Gianfranco Boccoli; Stefano Dellabella; Emilio Ceraudo; Stefano Santarelli
Journal:  Perit Dial Int       Date:  2015 Jul-Aug       Impact factor: 1.756

2.  Few Outflow Problems With a Self-locating Catheter for Peritoneal Dialysis: A Randomized Trial.

Authors:  Bernd G Stegmayr; Wolfgang Sperker; Christina H Nilsson; Christina Degerman; Sven-Erik Persson; Jan Stenbaek; Conny Arnerlöv
Journal:  Medicine (Baltimore)       Date:  2015-12       Impact factor: 1.817

  2 in total