Evidence for early loading of osseointegrated implants for bone conduction at 4 weeks.

Claire Marie McLarnon1, Ian Johnson, Tom Davison, John Hill, Beverly Henderson, Denise Leese, Suzanne Marley.   

Abstract

INTRODUCTION: The purpose of our study was to directly measure the stability of the abutment using resonance frequency analysis (RFA) and to report on the change in stability over time for implants loaded 4 weeks postoperatively.
MATERIALS AND METHODS: The principle behind RFA is to obtain a numerical value relating to stability. The Osstell RFA recording device measures the resonance in a magnetic field of a 1 cm commercially manufactured attachment (SmartPeg) that is screwed onto the bone conduction device abutment. RFA measurements were obtained at surgery, 1, 4, and 16 weeks after implant surgery. The degree of movement (vibration) is inversely proportional to the stability of the abutment, and a numerical figure, the implant stability quotient (ISQ), is derived. Patients were fitted with the new Cochlear Bone-Anchored Hearing Aid (BAHA) BI300 series implant using a 1-stage procedure and loaded 4 weeks postoperatively.
RESULTS: Prospectively 68 consecutive patients were recruited with a male to female ratio of 23:45. Nineteen patients had bilateral BAHA devices fitted, giving a total of 88 implants. The average ISQ change (delta) for the cohort gives a value of 0 at time of surgery. The change in ISQ at 1 week was -0.10 (95% confidence interval [CI], 0.74), at 4 weeks was 1.21 (95% CI, 0.59), and at 16 weeks was 1.60 (95% CI, 0.77). DISCUSSION: Successful early loading of the BAHA sound processor has been achieved. RFA provides a reliable means of assessing stability for loading and measuring implant stability in the longer term. These data support the evidence for early loading at 4 weeks with good clinical safety.

Entities:  

Mesh:

Year:  2012        PMID: 23069746     DOI: 10.1097/MAO.0b013e31826dba5f

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  9 in total

1.  Stability and survival of bone-anchored hearing aid implant systems in post-irradiated patients.

Authors:  Mark D Wilkie; Kathryn A Lightbody; Ali A Salamat; Kalyan M Chakravarthy; David A Luff; Robert H Temple
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-02-17       Impact factor: 2.503

2.  A new wide-diameter bone-anchored hearing implant-prospective 1-year data on complications, implant stability, and survival.

Authors:  Søren Foghsgaard; Per Caye-Thomasen
Journal:  Otol Neurotol       Date:  2014-08       Impact factor: 2.311

3.  Loading of osseointegrated implants for bone conduction hearing at 3 weeks: 3-year stability, survival, and tolerability.

Authors:  Rik C Nelissen; Christine A den Besten; Hubert T Faber; Catharina A J Dun; Emmanuel A M Mylanus; Myrthe K S Hol
Journal:  Eur Arch Otorhinolaryngol       Date:  2015-08-14       Impact factor: 2.503

4.  Stability, survival, and tolerability of a 4.5-mm-wide bone-anchored hearing implant: 6-month data from a randomized controlled clinical trial.

Authors:  Rik C Nelissen; Christine A den Besten; Emmanuel A M Mylanus; Myrthe K S Hol
Journal:  Eur Arch Otorhinolaryngol       Date:  2015-03-20       Impact factor: 2.503

5.  Stability, Survival, and Tolerability of an Auditory Osseointegrated Implant for Bone Conduction Hearing: Long-Term Follow-Up of a Randomized Controlled Trial.

Authors:  Christine A den Besten; Joacim Stalfors; Stina Wigren; Johan Ivarsson Blechert; Mark Flynn; Måns Eeg-Olofsson; Rohini Aggarwal; Kevin Green; Rik C Nelissen; Emmanuel A M Mylanus; Myrthe K S Hol
Journal:  Otol Neurotol       Date:  2016-09       Impact factor: 2.311

6.  Three-year Outcomes of a Randomized Controlled Trial Comparing a 4.5-mm-Wide to a 3.75-mm-Wide Titanium Implant for Bone Conduction Hearing.

Authors:  Ivo J Kruyt; Rik C Nelissen; Emmanuel A M Mylanus; Myrthe K S Hol
Journal:  Otol Neurotol       Date:  2018-06       Impact factor: 2.311

7.  Clinical Evaluation of a Novel Laser-Ablated Titanium Implant System for Bone Anchored Hearing Systems in a Pediatric Population and the Relationship of Resonance Frequency Analysis With Implant Survival.

Authors:  Max Sallis Osborne; Anne Child-Hymas; Marcus Holmberg; Peter Thomsen; Martin L Johansson; Ann-Louise McDermott
Journal:  Otol Neurotol       Date:  2022-02-01       Impact factor: 2.311

8.  Stability Testing of a Wide Bone-Anchored Device after Surgery without Skin Thinning.

Authors:  Malou Hultcrantz
Journal:  Biomed Res Int       Date:  2015-07-05       Impact factor: 3.411

9.  Clinical Outcome of a Wide-diameter Bone-anchored Hearing Implant and a Surgical Technique With Tissue Preservation.

Authors:  Marius S Mowinckel; Martin N Møller; Kirsten N Wielandt; Søren Foghsgaard