Literature DB >> 23068570

Multinational evaluation of the interpretability of the iterative method of optimisation of AV delay for CRT.

Claire E Raphael1, Andreas Kyriacou, Siana Jones, Punam Pabari, Graham Cole, Resham Baruah, Alun D Hughes, Darrel P Francis.   

Abstract

BACKGROUND: AV delay optimisation of biventricular pacing devices (cardiac resynchronisation therapy, CRT) is performed in trials and recommended by current guidelines. The Doppler echocardiographic iterative method is the most commonly recommended. Yet whether it can be executed reliably has never been tested formally.
METHODS: 36 multinational specialists, familiar with using the echocardiographic iterative method of CRT optimisation, were shown 20-40 sets of transmitral Doppler traces at 6-8 AV settings and asked to select the optimal AV delay. Unknown to the specialists, some Doppler datasets appeared in duplicate, allowing assessment of both inter and intra-specialist interpretation.
RESULTS: On the Kappa scale of agreement (1 = perfect agreement, 0 = chance alone), the agreement regarding optimal AV delay between specialists was poor (kappa=0.12 ± 0.08). More importantly, agreement of specialists with themselves (i.e. viewing identical sets of traces, twice) was also poor, with Kappa=0.23 ± 0.07 and mean absolute difference in optimum AV delay of 83 ms between first and second viewing of the same traces.
CONCLUSION: Iterative AV optimisation is not executed reliably by experts, even in an artificially simplified context where biological variability and variation in image acquisition are eliminated by use of identical traces. This cannot be blamed on insufficient skills of some experts or discordant methods of selecting the optimum, because operators also showed poor agreement with themselves when assessing the same trace. Instead, guidelines should retract any recommendation for this algorithm. Guideline-development processes might usefully begin with a rudimentary check on proposed algorithms, to establish at least minimal credibility.
Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  AV delay; AV optimisation; Atrioventricular delay; Biventricular pacemaker; CRT; Cardiac resynchronisation; Cardiac resynchronisation therapy; Iterative; LV; Left ventricle; SMART-AV

Mesh:

Year:  2012        PMID: 23068570     DOI: 10.1016/j.ijcard.2012.09.097

Source DB:  PubMed          Journal:  Int J Cardiol        ISSN: 0167-5273            Impact factor:   4.164


  3 in total

1.  Comparison of Echocardiographic and Electrocardiographic Mapping for Cardiac Resynchronisation Therapy Optimisation.

Authors:  Helder Pereira; Tom A Jackson; Simon Claridge; Jonathan M Behar; Cheng Yao; Benjamin Sieniewicz; Justin Gould; Bradley Porter; Baldeep Sidhu; Jaswinder Gill; Steven Niederer; Christopher A Rinaldi
Journal:  Cardiol Res Pract       Date:  2019-02-21       Impact factor: 1.866

Review 2.  Translating stem cell research to cardiac disease therapies: pitfalls and prospects for improvement.

Authors:  Michael R Rosen; Robert J Myerburg; Darrel P Francis; Graham D Cole; Eduardo Marbán
Journal:  J Am Coll Cardiol       Date:  2014-09-02       Impact factor: 24.094

3.  Multicenter Randomized Controlled Crossover Trial Comparing Hemodynamic Optimization Against Echocardiographic Optimization of AV and VV Delay of Cardiac Resynchronization Therapy: The BRAVO Trial.

Authors:  Zachary I Whinnett; S M Afzal Sohaib; Mark Mason; Edward Duncan; Mark Tanner; David Lefroy; Mohamed Al-Obaidi; Sue Ellery; Francisco Leyva-Leon; Tim Betts; Mark Dayer; Paul Foley; Jon Swinburn; Martin Thomas; Raj Khiani; Tom Wong; Zaheer Yousef; Dominic Rogers; Paul R Kalra; Vignesh Dhileepan; Katherine March; James Howard; Andreas Kyriacou; Jamil Mayet; Prapa Kanagaratnam; Michael Frenneaux; Alun D Hughes; Darrel P Francis
Journal:  JACC Cardiovasc Imaging       Date:  2018-05-16
  3 in total