Venobronchial fistula: a complication associated with central venous catheterization for chemotherapy.

J F Beauregard1, A H Matsumoto, M G Paul, R W Holt.   

Abstract

We report a case in which a venobronchial fistula developed 4 months after insertion of a silicone rubber chronic central venous catheter for chemotherapy administration. In retrospect, the tip of the catheter was in the azygous vein rather than in the superior vena cava. The suboptimal position of the catheter tip, in combination with the infusion of sclerosing chemotherapeutic agents and hyperalimentation solution, may have predisposed the patient to this complication. Accurate localization of the tip of central venous catheters at the time of placement should minimize catheter-related complications, including the rare complication of venobronchial fistula. After placement of chronic central venous catheters, biplanar chest X-rays should be obtained to determine catheter tip location prior to the instillation of chemotherapeutic agents. Any residual questions concerning the catheter tip location after biplanar X-rays should be addressed by prompt venography through the catheter.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2306768     DOI: 10.1002/ccd.1810190114

Source DB:  PubMed          Journal:  Cathet Cardiovasc Diagn        ISSN: 0098-6569


  2 in total

1.  Venobronchial fistula: an unusual early complication after central venous catheterization.

Authors:  V Guariento; G B Luciani; B Dal Corso; A Carletti; A Mazzucco; A Luzzani
Journal:  Intensive Care Med       Date:  1997-11       Impact factor: 17.440

2.  Leaking stent leading to a venous-bronchial fistula.

Authors:  Raghav Gupta; Andre Jared; James Mahoney
Journal:  Int J Crit Illn Inj Sci       Date:  2017 Oct-Dec