Clinical evaluation of a novel test strip technology for blood glucose monitoring: accuracy at hypoglycaemic glucose levels.

Srikanth Bellary1, Hillary Cameron, Kirsty Macleod, Michael Malecha, Krisna Koria, Praveen Raja, Jesús Diago Cabezudo, John Ellison.   

Abstract

AIM: To evaluate OneTouch® Verio™ test strip performance at hypoglycaemic blood glucose (BG) levels (<3.9 mmol/L [<70 mg/dL]) at seven clinical studies.
METHODS: Trained clinical staff performed duplicate capillary BG monitoring system tests on 700 individuals with type 1 and type 2 diabetes using blood from a single fingerstick lancing. BG reference values were obtained using a YSI 2300 STAT™ Glucose Analyzer. The number and percentage of BG values within ±0.83 mmol/L (±15 mg/dL) and ±0.56 mmol/L (±10 mg/dL) were calculated at BG concentrations of <3.9 mmol/L (<70 mg/dL), <3.3 mmol/L (<60 mg/dL), and <2.8 mmol/L (<50 mg/dL).
RESULTS: At BG concentrations <3.9 mmol/L (<70 mg/dL), 674/674 (100%) of meter results were within ±0.83 mmol/L (±15 mg/dL) and 666/674 (98.8%) were within ±0.56 mmol/L (±10 mg/dL) of reference values. At BG concentrations <3.3 mmol/L (<60 mg/dL), and <2.8 mmol/L (<50 mg/dL), 358/358 (100%) and 270/270 (100%) were within ±0.56 mmol/L (±10 mg/dL) of reference values, respectively.
CONCLUSION: In this analysis of data from seven independent studies, OneTouch Verio test strips provide highly accurate results at hypoglycaemic BG levels.
Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23063304     DOI: 10.1016/j.diabres.2012.09.027

Source DB:  PubMed          Journal:  Diabetes Res Clin Pract        ISSN: 0168-8227            Impact factor:   5.602


  3 in total

1.  Seven-Year Surveillance of the Clinical Performance of a Blood Glucose Test Strip Product.

Authors:  Steven Setford; Mike Grady; Stuart Phillips; Lesley Miller; Stephen Mackintosh; Hilary Cameron; Krisna Corrigall
Journal:  J Diabetes Sci Technol       Date:  2017-04-13

2.  In response to Teodorczyk and coauthors: System accuracy of blood glucose monitoring devices according to the current and proposed ISO 15197 standards.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Stefan Pleus; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

3.  System accuracy of blood glucose monitoring devices according to the current and proposed ISO 15197 standards.

Authors:  Maria Teodorczyk; Santhanagopalan Nandagopalan; Patricia Maguire; Janet Stegmann
Journal:  J Diabetes Sci Technol       Date:  2013-05-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.