Analysis of the bacterial seal at the implant-abutment interface in external-hexagon and Morse taper-connection implants: an in vitro study using a new methodology.

Mario Eduardo Jaworski1, Ana Cláudia Moreira Melo, Cyntia Maria Telles Picheth, Ivete Aparecida de Mattias Sartori.   

Abstract

PURPOSE: The aim of this study was to compare, in vitro, external-hexagon and Morse taper implant systems with respect to bacterial sealing between implants and abutments using a new methodology.
MATERIALS AND METHODS: Two groups of implants were tested. Group 1 implants had an external-hexagon implant-abutment interface (Neodent) and group 2 implants featured a Morse taper (Neodent) interface. The implants were perforated apically with a 1-mm bur until the bur reached the internal chamber. Prosthetic components were adapted with the recommended torque (32 and 10 Ncm, respectively) for each group. The implants were attached to an assay vial, with the abutment end positioned into the tube. With a sterilized syringe, the assay vials were filled with liquid culture medium (brain heart infusion broth). All the specimens were sterilized by gamma radiation. After the efficacy of sterilization had been confirmed using control samples, the apical hole was carefully opened and inoculated with Escherichia coli.
RESULTS: Samples were examined daily for evidence of contamination. Within a 14-day period, 60% of the samples of group 1 were contaminated and 30% of group 2 samples were contaminated. After this period there was no further contamination in either group.
CONCLUSION: Although both systems exhibited bacterial contamination, the Morse taper implants of the system used in this study provided a better bacterial seal than external-hexagon implants of the same system.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23057021

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  6 in total

1.  Bacterial contamination along implant-abutment interface in external and internal-hex dental implants.

Authors:  Greison Rabelo de Oliveira; Sergio Olate; Leandro Pozzer; Lucas Cavalieri-Pereira; Jaime G Rodrigues-Chessa; José Ricardo Albergaría-Barbosa
Journal:  Int J Clin Exp Med       Date:  2014-03-15

2.  Determination of fluid leakages in the different screw-retained implant-abutment connections in a mechanical artificial mouth.

Authors:  D Martin-Gili; M Molmeneu; M Fernandez; M Punset; Ll Giner; J Armengou; F Javier Gil
Journal:  J Mater Sci Mater Med       Date:  2015-07-15       Impact factor: 3.896

Review 3.  Is the internal connection more efficient than external connection in mechanical, biological, and esthetical point of views? A systematic review.

Authors:  Marcelo Coelho Goiato; Eduardo Piza Pellizzer; Emily Vivianne Freitas da Silva; Liliane da Rocha Bonatto; Daniela Micheline dos Santos
Journal:  Oral Maxillofac Surg       Date:  2015-04-25

4.  A new system of implant abutment connection: how to improve a two piece implant system sealing.

Authors:  F Grecchi; M DI Girolamo; F Cura; V Candotto; F Carinci
Journal:  Oral Implantol (Rome)       Date:  2017-11-30

5.  Bacterial infiltration and detorque at the implant abutment morse taper interface after masticatory simulation.

Authors:  Ana Paula Granja Scarabel Nogueira Bella; Alessandra Sayuri Tuzita; Ivana Barbosa Suffredini; Alberto Noriyuki Kojima; Elcio Magdalena Giovani; Alfredo Mikail Melo Mesquita
Journal:  Sci Rep       Date:  2022-10-12       Impact factor: 4.996

6.  Comparative evaluation of peri-implant tissues in patients wearing mandibular overdenture with different implant platforms.

Authors:  Laércio Almeida de Melo; Danielle Bezerra de Farias; Annie Karoline Bezerra de Medeiros; Gustavo Augusto Seabra Barbosa; Euler Maciel Dantas; Adriana da Fonte Porto Carreiro
Journal:  J Indian Soc Periodontol       Date:  2017 Nov-Dec
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.