Taro Kimura1, Yoshimasa Yokoyama, Daisuke Sakota, Eiki Nagaoka, Takashi Kitao, Kazuo Takakuda, Setsuo Takatani.
Abstract
The impact of continuous flow left ventricular assist device (LVAD) pumping on platelet aggregation was investigated in animal experiments utilizing six calves. A single-use MagLev centrifugal blood pump, MedTech MagLev, was used to bypass the calves' hearts from the left atrium to the descending aorta at a flow rate of 50 ml/kg/min. The LVAD's impact on blood coagulation activities was evaluated based on the platelet aggregability, which was measured with a turbidimetric assay method during the preoperative, operative, and postoperative periods. Heparin and warfarin were used for anticoagulation, while aspirin was used for the antiplatelet therapy. A decrease in platelet aggregation immediately after the pump started was observed in the cases of successful long-term pump operation, while the absence of such a decrease might have caused coagulation-related complications to terminate the experiments. Thus, the platelet aggregability was found to be significantly affected by the pump, and its initial trend may be related to the long-term outcome of the mechanical circulatory support.Entities:
Mesh:
Substances:
Year: 2012 PMID: 23053045 DOI: 10.1007/s10047-012-0664-2